Figure 4-4 Voltage and Current Limit Settings
4.2.3 Setting the Output Voltage and Current Limit
Set the output voltage and current limit by highlighting the appropriate value using the
left and right arrow keys, adjusting the value using the up down arrow keys, pressing
ENT
again to save to value. Use the arrow keys to move to the next value, or to the
Back
or
Next
buttons as required.
Select the output voltage to best suit the voltage of the system under test. For example,
if the system operates at 100V DC, choose an output voltage of 110V. Settings of 24V,
48V, 110V, 220V, 500V, and 1000V are available.
The current limit should be used only where high levels of current may cause alarms or
protection systems to be triggered. For example, railway signaling systems may indicate
an alarm if leakage current
– or injected current from the signal generator in this case –
exceeds 5mA. In this case the current limit should be set to 5mA. If no earth fault