Recording Mode
Record Until Memory Used (Stop recording when full) or Continuous FIFO
(First In - First Out - overwrites oldest data when full).
Recording Sample Interval
From: Every 1; 2; 5; 10; 15; 30 Seconds, or Every 1; 2; 5; 10; 15; 30 Minutes.
Recorder Trigger
The recording Start/Stop trigger method. From: Operation Mode; Recorder Menu;
On Alarm; Digital Input
Trigger On Alarms
Any from: Alarm n – Where n is alarms 1 to 5. Any combination of these can be set
to trigger (TRG) or not (OFF).
Values To Record
Any from: Process Variable value; Maximum or Minimum PV (since previous
sample); Setpoint; Primary Power or Secondary Power.
Events To Record
Any from: Alarm n Status or Unit On/Off. Note: An alarm state change between
samples is also recorded. This uses additional recorder memory.
Recorder Status Information
Shows if a recording is in progress; the recording mode; memory usage per sample;
memory remaining and approximate recording time remaining.
Recorder Clock Configuration:
Refer to the Supplementary Product Manual for information about the additional screens when Data
Recorder is fitted.
Date Format
w The format used for displayed dates: dd/mm/yyyy (Day / Month / Year) or mm/dd/
yyyy (Month / Day / Year). – Recorder versions only.
Set Date
w Sets the internal clock Date. – Entered in the format defined by Date Format screen.
– Recorder versions only.
Set Day Of Week
w Sets the day of week used by the internal clock. – Recorder versions only.
Set Time
w Sets the internal clock Time. - In hh:mm:ss (Hours : Minutes : Seconds) format.
– Recorder versions only.
Display Configuration:
Enable Custom Display Mode
Enables/disables Custom Operation Mode, if configured
(requires PC configuration software).
Read Only Operation Mode?
Allows Operation Mode to be Read-Only or Read/Write. Screens can be seen but,
values cannot be changed if Read-Only.
Operation Mode Bar Graph Format From: PID Power or Control Deviation or.
Trend Sample Interval
Interval between display of next value on the trend graph From: Every 1; 2; 5; 10; 15;
30 Seconds, or Every 1; 2; 5; 10; 15; 30 Minutes.
Select Trend Mode
From: PV only, PV and SP or Max/Min PV between samples (candle-stick graph).
Alarm activity is always shown.
Display Colour
From: Red only; Green only; Red to Green on Alarm or Green to Red on Alarm.
Invert Display
Standard or Negative display image.
Display Contrast
Screen contrast (0 and 100) to improve clarity. 100 = maximum contrast.
Language
Select English or the alternate local language. The alternate language type can be
changed using the PC software.
Lock Code Configuration:
Lock Code View 1
View and edit the Setup Wizard; Configuration Mode; Tuning Menu, Supervisor
Mode, USB Menu and Recorder menu Lock Codes (1-9999 or OFF).
Default Values = 10
Reset To Defaults:
Reset To Defaults
Set all parameters to default values. Caution: User must reconfigure all required
settings before using the instrument following a reset.
*Both Recorder Trigger state and Digital input selection must be the same to start recording
** Applies to Temperature control Only not pressure control.