From lab to production,
providing a window into the process
-12-
www.dynisco.com
Rev: n/a
P/N: n/a
ECO: n/a
3. Power Up Sequence
Following the power-up self-test and logo screen, the instrument normally enters Operation Mode,
from which the user can select the instrument’s Main Menu (refer to the Screen Sequence list). The
exceptions to this are the first power-up after purchase, when option modules have been changed
or if an error is detected.
Setup Wizard
An easy Setup Wizard runs automatically at first ever power-up. Follow the Wizard to setup
parameters required for typical applications (screens marked w in the Screen Sequence list). A
partial Wizard also runs whenever option modules have been changed, this only shows
parameters affected by the changes. The Wizard can also be run from the Main Menu. It exits to
Operation Mode once completed.
Start-Up Errors
These messages indicate that a hardware or configuration an error has occurred.
Caution: Do not continue with the process until the issue is resolved.
Message Displayed
Reason
Option Slot
n
Error
Fault detected. Replace the module in slot
n
Configuration Problem
Check all instrument parameters before using
For Service Contact
Details of who to contact if a fault persists
4. Operation Mode
This mode is entered at power on, or accessed from the Main Menu. If required, all Operation Mode
parameters can be made read only
(see Display Configuration).
Note: Configuration must be completed before starting normal operations.
Normal Operation
Subsequent screens allow the display and selection/adjustment* alarm status and trends.
LED Indicators
LED Function Labels
Engineering Units
Input as Percentage of Span
Process Variable Value 1
Process Variable Value 2
Typical Operation Screen