From lab to production,
providing a window into the process
-7-
www.dynisco.com
Rev: n/a
P/N: n/a
ECO: n/a
Strain Gauge Inputs: 350 to 5K Strain Gage
Bridge Connection 4 or 6 wire (6 to use internal shunt cal switch)
Bridge Excitation 10 V +/- 7%
Bridge Sensitivity 1.4 - 4 mV/V
Input Signal Span - 25% to +125% of full scale (approximately - 10 mV
to +50 mV)
Calibration Internal switch between CAL2 & CAL1 terminals(7&6 or 39&40).
External resistor only.
Shunt Value From 40% to 100%
Supported
Thermocouple
Types & Ranges:
Thermocouple
Calibration:
Supported RTD
Types & Ranges:
RTD Calibration:
RTD Excitation:
Sensor current 150 A ± 10%.
Type
Range °C
Range °F
B
+100 to 1824°C
+211 to 3315°F
C
0 to 2320°C
32 to 4208°F
D
0 to 2315°C
32 to 4199°F
E
-240 to 1000°C
-400 to 1832°F
J
-200 to 1200°C
-328 to 2192°F *
K
-240 to 1373°C
-400 to 2503°F *
L
0 to 762°C
32 to 1402°F *
N
0 to 1399°C
32 to 2551°F *
PtRh 20%:40%
0 to 1850°C
32 to 3362°F
R
0 to 1759°C
32 to 3198°F
S
0 to 1762°C
32 to 3204°F
T
-240 to 400°C
-400 to 752°F *
Optional decimal place can be displayed up to 999.9°C/F
± 0.1% of full range, ± 1LSD (± 1°C for internal CJC if enabled). Linearization
better than better ± 0.2°C (± 0.05 typical) on ranges marked * in the table
above. Linearization for other ranges is better than better than ± 0.5°C.
BS4937, NBS125 & IEC584
Type
Range °C
Range °F
3-Wire PT100 -199 to 800°C
-328 to 1472°F
NI120
-80 to 240°C
-112 to 464°F
Optional decimal place can be displayed up to 999.9°C/F
0.1% of full range,
±
1LSD.
Linearization better than
±
0.2
°
C (
±
0.05 typical).
PT100 input to BS1904 & DIN43760 (0.00385 / /°C).