Electrotherapy Usage Cautions
The following are general cautions to be observed during interferential, premodulated, Russian,
biphasic, and high voltage pulsed stimulation. For microcurrent electrotherapy, see separate
“Microcurrent Usage Cautions” following this section.
Electrical stimulation, by its very nature, has the ability to irritate the patient’s skin. Certain
precautions should be observed to assure maximum safety and comfort for patients. A
patient’s tendency to have adverse reactions is dependent upon several factors. These factors
are:
Current Density. This is the amount of current being delivered to the patient divided by the
area through which the current is being delivered (the surface area of the electrodes being
used).
Electrode Condition. Worn or dried out electrodes cause the current to concentrate in small
areas of the electrode instead of going into the skin distributed evenly throughout the entire
electrode surface. This has the effect of increasing the current density, since the current is
being delivered through a smaller area.
Patient Susceptibility. Some patients’ skin is more sensitive to electrotherapy currents.
This can cause a reaction similar to a heat rash.
Electrotherapy treatment can result in a rash, burn, or blister. The tendency to do this is
dependent upon the factors listed above and can be minimized by applying the following
guidelines:
1. Use only moderate current. It is not always necessary to raise the treatment intensity to
just short of the patient’s pain threshold to achieve adequate results. Suggested
maximum treatment levels for different electrode sizes are given below (for interferential
or premodulated therapy). These values are not intended to be inflexible, but they should
be a guideline. If your treatment goes beyond these intensities, you should examine the
treatment to discover why such a high intensity setting is required.
Use as large an electrode as is practical for the application. Note that the current density in a
1.25” square electrode is FIVE TIMES the current density in a 1.75” by 3.75” electrode for
the same intensity setting. Using larger electrodes allows current to be delivered over a
larger area of the body, keeping the current density to be as low as possible and minimizing
the possibility for adverse reactions.
Interferential / Premodulated
Electrode
Size Maximum Recommended Intensity
Carbon Electrodes
3” round
25-30
3” x 5”
30-40
Self-adhesive Electrodes
1.75” square
10-15
1.75” x 3.75”
25-30
1.25” round
10-12
2” round
10-20
3” round
25-30
Summary of Contents for Dynatron 550 Plus
Page 1: ...i SERVICE MANUAL Dynatron 850plus Dynatron 550plus ...
Page 9: ...Dynatron 850plus 550plus 1 Section I Introduction ...
Page 106: ...Dynatron 850plus 550plus Ultrasound Technical Information 98 ...
Page 120: ...Dynatron 850plus 550plus Ultrasound Technical Information 112 ...
Page 121: ...Dynatron 850plus 550plus Service 113 Section VI Technical Information ...
Page 130: ...Dynatron 850plus 550plus General Technical Information 122 ...
Page 131: ...Dynatron 850plus 550plus Service 123 Section VII Service ...
Page 139: ...131 ...
Page 140: ...132 ...
Page 141: ...133 ...
Page 142: ...134 ...
Page 143: ...135 ...
Page 144: ...136 ...
Page 145: ...137 ...
Page 146: ...138 ...
Page 147: ...139 ...
Page 148: ...140 ...
Page 149: ...141 ...
Page 150: ...142 ...
Page 151: ...143 ...
Page 152: ...144 ...
Page 153: ...145 ...