High Volt Instructions
WARNING
• NEVER turn the power on or off while the unit is connected to the patient.
• Always give the Remote Stop device to the patient prior to the treatment.
• Always STOP a treatment before removing or attaching electrodes or leads. Leads and
electrodes must only be applied to the patient before a treatment is started.
• Never use worn or damaged leads or electrodes as these may result in injury to the
patient.
• See “Contraindications, Warnings, and Precautions” in this manual before administering a
treatment to a patient.
High volt electrical stimulation is a pulsed DC current with pulse widths in the microsecond
range and pulse rates ranging from 1 to 200 Hz, with a peak amplitude of up to 1.0 amp. The
Dynatron 850plus and 550plus deliver high volt utilizing a twin-peak monophasic waveform.
High volt treatments with the Dynatron
550plus and 850plus may be delivered using
electrodes or an optional hand-held probe.
The device provides a dedicated channel for
high volt treatment (HV). This channel is
used to deliver high volt treatments only.
The device’s other output channels (CH1-2)
remain available for other stim treatments
simultaneously with the high volt treatment.
Set Up High Volt Treatment with
Electrodes
This treatment setup utilizes a double-output
lead with two electrodes; an active and a
dispersive electrode. The size of the dispersive electrode is recommended to be double the
area of the active electrode. If desired, the active output of the lead may be bifurcated to
attach additional active electrodes; however, the combined total area of the active electrodes
should be no more than half the area of the single passive electrode. These setups are
illustrated on this page and the following page.
DOUBLE
OUTPUT LEAD:
Connect to "HV"
channel
Dispersive electrode:
Attach to NEGATIVE
lead output.
Active electrode:
Attach to POSITIVE
lead output.
( – )
( + )
High Volt Electrode Setup
Summary of Contents for Dynatron 550 Plus
Page 1: ...i SERVICE MANUAL Dynatron 850plus Dynatron 550plus ...
Page 9: ...Dynatron 850plus 550plus 1 Section I Introduction ...
Page 106: ...Dynatron 850plus 550plus Ultrasound Technical Information 98 ...
Page 120: ...Dynatron 850plus 550plus Ultrasound Technical Information 112 ...
Page 121: ...Dynatron 850plus 550plus Service 113 Section VI Technical Information ...
Page 130: ...Dynatron 850plus 550plus General Technical Information 122 ...
Page 131: ...Dynatron 850plus 550plus Service 123 Section VII Service ...
Page 139: ...131 ...
Page 140: ...132 ...
Page 141: ...133 ...
Page 142: ...134 ...
Page 143: ...135 ...
Page 144: ...136 ...
Page 145: ...137 ...
Page 146: ...138 ...
Page 147: ...139 ...
Page 148: ...140 ...
Page 149: ...141 ...
Page 150: ...142 ...
Page 151: ...143 ...
Page 152: ...144 ...
Page 153: ...145 ...