Dynatron
®
850plus & 550plus
Simultaneous Treatments
62
Simultaneous Treatments
The Dynatron 850plus and 550plus allow several combinations of simultaneous treatments to
be delivered at once using available channels. Simultaneous treatments are not the same as
combo treatments. A combo treatment combines ultrasound with a stim therapy into a single
treatment. A combo treatment is always delivered to one patient. Simultaneous treatments
are independent treatments that are set up separately, that have separate treatment timers, and
which may be delivered to one or more patients at the same time.
You can set up any number of separate treatments as described below with the exceptions
noted:
•
Channels 1 and 2 may be used for one interferential, two premodulated, one
premodulated and one Russian, or one premodulated and one biphasic treatment. For
treatments using one channel, the device will select the next available channel. For
treatments using two channels, the device will select both channels 1 and 2.
•
Russian or biphasic parameters selected will apply to both channels in a channels 1 and
2 at once. Therefore, you cannot have both a Russian and a biphasic treatment
simultaneously. You may deliver two biphasic or two Russian treatments.
• A
separate
High Volt treatments may be set up regardless of any other treatments that
may be currently running.
• A
separate
Ultrasound treatment may be set up, or a combination treatment using
ultrasound and a single-channel modality with any available channel.
•
The device allows only ONE attended therapy to be delivered at one time
. Attended
therapies include ultrasound, microcurrent probe, and high volt probe treatments. While
the attended therapy is in progress, other available channels may be used for unattended
treatments (subject to availability of the channel needed).
•
Microcurrent therapy is delivered only through Channel 1 and through the
Microcurrent Probes output jack at the same time. Whenever any microcurrent
treatment is in progress (probes or electrodes), Channel 1 is unavailable for any other use.
However, Channel 2 may be used for other stim treatments.
Set Up A Second Treatment
To set up the second (or third) treatment, after you have set up and started the first modality,
press the modality key for the second treatment to be set up. The device automatically selects
the treatment channel(s) to be used. The green channel light shows you the channel(s)
selected for this treatment. Plug the lead or cable into the corresponding output jack(s) before
you proceed with setting up the treatment.
Select the treatment parameters for the second treatment following the setup instructions for
that modality provided in this manual. When parameters have been entered, press START.
Summary of Contents for Dynatron 550 Plus
Page 1: ...i SERVICE MANUAL Dynatron 850plus Dynatron 550plus ...
Page 9: ...Dynatron 850plus 550plus 1 Section I Introduction ...
Page 106: ...Dynatron 850plus 550plus Ultrasound Technical Information 98 ...
Page 120: ...Dynatron 850plus 550plus Ultrasound Technical Information 112 ...
Page 121: ...Dynatron 850plus 550plus Service 113 Section VI Technical Information ...
Page 130: ...Dynatron 850plus 550plus General Technical Information 122 ...
Page 131: ...Dynatron 850plus 550plus Service 123 Section VII Service ...
Page 139: ...131 ...
Page 140: ...132 ...
Page 141: ...133 ...
Page 142: ...134 ...
Page 143: ...135 ...
Page 144: ...136 ...
Page 145: ...137 ...
Page 146: ...138 ...
Page 147: ...139 ...
Page 148: ...140 ...
Page 149: ...141 ...
Page 150: ...142 ...
Page 151: ...143 ...
Page 152: ...144 ...
Page 153: ...145 ...