Dynatron
®
850plus & 550plus
ULTRASOUND - Treatment Instructions
52
Poor Coupling
If all or part of the soundhead is not touching the patient, coupling may not be adequate and
the patient will not receive the intended dosage of the ultrasonic wave. Poor coupling, if
allowed to continue, can also result in increased temperature of the soundhead. Do not hold
the soundhead in the air while a treatment is running as this may damage the soundhead
crystal and/or stress electronic components in the device.
Ultrasound Coupling Bar Graph
This feature for models having software greater than Rev. 1.01 only (see Page 2).
During an ultrasound treatment, you can press SOUND-TARGET (press both keys at once)
to display a bar graph to indicate coupling. Coupling sensing indicates whether adequate
contact is being maintained between the soundhead and the patient’s skin. The bar graph
uses the eight segments in the top half of the display, with eight lighted segments indicating
best conductance, and one lighted segment indicating poorest conductance. In the illustration
below, eight of the eight segments are lighted indicating good coupling.
POWER-INTENSITY
TIME
START
STOP
PAUSE
WIDTH
RATE
FREQ
TIME
SOUND
MICRO
HV 1
2
1
FUNCTION
Press the SOUND and the TARGET keys
simultaneously during an ultrasound treatment to
view a graph showing soundhead coupling. Press
Time to return to the treatment time display.
CH
Ultrasound Coupling Bar Graph
PREMOD
IFC
HIGH
LOW
TARGET
SWEEP
COMBO
SOUND
1 MHz
2 MHz
3 MHz
10%
20%
50%
CONT
Coupling Bar Graph. Eight lighted segments indicate good coupling between patient and soundhead.
To correct poor coupling ensure you are using an adequate amount of ultrasound conductive
gel, and the face of the soundhead is making full contact with the patient’s skin.
Press TIME to redisplay the treatment time. Treatment time continues counting down while
the bar graph is displayed unless the treatment is paused for any reason.
Disable the Coupling Sensing. You can disable the coupling sensing capability of the
device if desired. While the treatment is running, press the FUNCTION and the
Power/Intensity DOWN ARROW keys at the same time. The Time display will very briefly
display the new setting as follows:
CP 1 = Coupling on
CP 0 = Coupling off
Summary of Contents for Dynatron 550 Plus
Page 1: ...i SERVICE MANUAL Dynatron 850plus Dynatron 550plus ...
Page 9: ...Dynatron 850plus 550plus 1 Section I Introduction ...
Page 106: ...Dynatron 850plus 550plus Ultrasound Technical Information 98 ...
Page 120: ...Dynatron 850plus 550plus Ultrasound Technical Information 112 ...
Page 121: ...Dynatron 850plus 550plus Service 113 Section VI Technical Information ...
Page 130: ...Dynatron 850plus 550plus General Technical Information 122 ...
Page 131: ...Dynatron 850plus 550plus Service 123 Section VII Service ...
Page 139: ...131 ...
Page 140: ...132 ...
Page 141: ...133 ...
Page 142: ...134 ...
Page 143: ...135 ...
Page 144: ...136 ...
Page 145: ...137 ...
Page 146: ...138 ...
Page 147: ...139 ...
Page 148: ...140 ...
Page 149: ...141 ...
Page 150: ...142 ...
Page 151: ...143 ...
Page 152: ...144 ...
Page 153: ...145 ...