Dynatron
®
850plus & 550plus
Installation & Features
15
Time and Power Displays
The “Time” display and the “Power-Intensity” display can show the settings for only one
channel at a time. The Time and Power-Intensity settings displayed are for the channel with
the GREEN light only. Any other channel in use at that time will have a YELLOW light to
show it is active but its parameters are not currently displayed.
To view the settings for another
channel or output, press the channel
toggle key one or more times until
the light for the desired channel
becomes GREEN. The GREEN
light appears next to a different
channel or output each time you
press the toggle key, and the Time
and Power-Intensity displays change
to show the parameters currently in
effect for that channel.
The green and yellow channel lights
each will also appear solid (non-
flashing) or flashing. A solid light
means current is being delivered to
the channel at this time (for example, during the ON cycle of the Russian stimulation
treatment). A flashing light means current is not being delivered to the channel at this time
(for example, during the OFF cycle of the Russian stimulation treatment).
CHANNEL / OUTPUT INDICATOR LIGHTS
GREEN
Solid •
You
CAN see this channel’s parameters displayed on Time
and Power-Intensity displays.
•
The
channel
IS delivering current.
GREEN
Flashing •
You
CAN see this channel’s parameters displayed on Time
and Power-Intensity displays.
•
The
channel
IS NOT delivering current.
YELLOW
Solid •
You
CANNOT see this channel’s parameters displayed on
Time and Power-Intensity displays.
•
The
channel
IS delivering current.
YELLOW
Flashing •
You
CANNOT see this channel’s parameters displayed on
Time and Power-Intensity displays.
•
The
channel
IS NOT delivering current.
Ultrasound Heads (Dynatron 850plus only)
The Dynatron 850plus features the new Dynatronics SmartHead™ ultrasound applicator
soundhead that has all calibration information self-contained. SmartHeads may be used
interchangeably with any 50 Series Plus model or the Dynatron 125 model with no need for
entering calibration numbers into the device.
TIME
WIDTH
RATE
FREQ
TIME
SOUND
MICRO
HV 1
2
1
Press the
channel toggle
key to select any
channel that is
currently in use.
The LED for the channel selected becomes GREEN and the
LEDs for all other active channels become YELLOW.
When a channel is selected its parameters are displayed in
the Time and Power-Intensity displays, and treatment
parameters for that treatment may be modified.
CH
Channel Toggle Key
Summary of Contents for Dynatron 550 Plus
Page 1: ...i SERVICE MANUAL Dynatron 850plus Dynatron 550plus ...
Page 9: ...Dynatron 850plus 550plus 1 Section I Introduction ...
Page 106: ...Dynatron 850plus 550plus Ultrasound Technical Information 98 ...
Page 120: ...Dynatron 850plus 550plus Ultrasound Technical Information 112 ...
Page 121: ...Dynatron 850plus 550plus Service 113 Section VI Technical Information ...
Page 130: ...Dynatron 850plus 550plus General Technical Information 122 ...
Page 131: ...Dynatron 850plus 550plus Service 123 Section VII Service ...
Page 139: ...131 ...
Page 140: ...132 ...
Page 141: ...133 ...
Page 142: ...134 ...
Page 143: ...135 ...
Page 144: ...136 ...
Page 145: ...137 ...
Page 146: ...138 ...
Page 147: ...139 ...
Page 148: ...140 ...
Page 149: ...141 ...
Page 150: ...142 ...
Page 151: ...143 ...
Page 152: ...144 ...
Page 153: ...145 ...