Dynatron
®
850plus & 550plus
Ultrasound Technical Information
100
Ultrasound Regulation
The Dynatron 850plus complies with the following:
•
IEC 601-2-5: 1984. Medical Electrical Equipment Part 2: Particular requirements
for the safety of ultrasonic therapy equipment.
•
FDA 21CFR 1050(c)(1)(i). The error in indication of the temporal-average
ultrasonic power shall not exceed ±20 percent for all emissions greater than 10
percent of the maximum emission.
•
FDA 21CFR 1050(c)(1)(ii). The sum of the errors in the indications of temporal-
maximum ultrasonic power and the ratio of the temporal-maximum effective
intensity to the temporal-average effective intensity shall not exceed ±20 percent for
all emissions greater than 10 percent of the maximum emission.
•
FDA 21CFR 1050.10(c)(2). The treatment timer must be accurate to within 0.5
minute of the preset duration of emission for settings less than 5 minutes, to within
10 percent of the preset duration of emission for settings of from 5 minutes to 10
minutes, and to within 1 minute of the preset duration of emission for settings greater
than 10 minutes.
NOTE: The Dynatron 850plus is accurate to within ±1% of any treatment time.
Pursuant to FDA 21CFR 1050.10(f)(1), the uncertainties in magnitude, expressed in
percentage error, of the ultrasonic frequency, effective radiating area, and the ratio of the
temporal-maximum to temporal-average effective intensity, pulse duration, and pulse
repetition rate for the Dynatron 850plus are as follows:
(1) Ultrasonic
frequency ..................................... ±15%
(2)
Effective Radiating Area............................... ±50%
(3)
Ratio of the temporal-maximum to
temporal-average
effective intensity ............. ±20%
(4) Pulse
duration................................................ ±10%
(5)
Pulse repetition rate....................................... ±10%
Summary of Contents for Dynatron 550 Plus
Page 1: ...i SERVICE MANUAL Dynatron 850plus Dynatron 550plus ...
Page 9: ...Dynatron 850plus 550plus 1 Section I Introduction ...
Page 106: ...Dynatron 850plus 550plus Ultrasound Technical Information 98 ...
Page 120: ...Dynatron 850plus 550plus Ultrasound Technical Information 112 ...
Page 121: ...Dynatron 850plus 550plus Service 113 Section VI Technical Information ...
Page 130: ...Dynatron 850plus 550plus General Technical Information 122 ...
Page 131: ...Dynatron 850plus 550plus Service 123 Section VII Service ...
Page 139: ...131 ...
Page 140: ...132 ...
Page 141: ...133 ...
Page 142: ...134 ...
Page 143: ...135 ...
Page 144: ...136 ...
Page 145: ...137 ...
Page 146: ...138 ...
Page 147: ...139 ...
Page 148: ...140 ...
Page 149: ...141 ...
Page 150: ...142 ...
Page 151: ...143 ...
Page 152: ...144 ...
Page 153: ...145 ...