Usage
15 Maintenance
15.1 Changing the valve in the fluid
container
The valve on the underside of the fluid container
must be cleaned regularly and checked for
blockages and leaks
i
Press the valve lightly against the fluid con-
tainer and unscrew it anti-clockwise
i
Clean the valve
If cleaning is not possible, eg if the filter in
the valve is blocked, then the valve needs to
be replaced
i
Check the O-ring
In the event of leaks, loose seating or visible
damage the O-ring must be replaced
i
Insert the valve in the holder and tighten it
clockwise as far as it will go
2. Decontamination process was not carried
out before surgery break:
i
Perform reprocessing before initial start-up of
the unit, see "13 Reprocessing" on page 29