![Durr Dental Vector Scaler Installation And Operating Instructions Manual Download Page 36](http://html.mh-extra.com/html/durr-dental/vector-scaler/vector-scaler_installation-and-operating-instructions-manual_2542917036.webp)
34
9000-615-32/02 1808V002
Usage
the required reaction time, so that all parts are
covered
i
Clean all accessible external surfaces and in-
ternal surfaces below the surface for 5 min-
utes with a sterile cleaning brush until all visi-
ble contaminants have been removed
i
Place instruments in small parts baskets into
the ultrasonic bath
i
Place the handpiece without covers in a con-
tainer with fluid The drive mechanism of the
handpiece must not lie in the fluid (malfunc-
tion) Therefore observe instructions concern-
ing maximum fluid level for the Scaler hand-
piece
max.
i
Place container with the handpieces in the ul-
trasonic bath with a suitable carrier
i
Rinse handpieces through at least 3 times us-
ing a 20-ml disposable pipette
20 ml
i
Remove all rinse adapters
i
Remove all rinse adapters
i
Screw Scaler instruments onto the rinsing
adapter one after another and rinse through
each of the instruments at least 3 times using
a 20-ml disposable syringe
20 ml
i
Comply with the cleaning agents' action times
specified by the manufacturer
Intermediate rinsing
After the action time prescribed by the manufac-
turer:
i
Rinse all components under running water for
at least 1 minute (temperature < 20 °C)
i
Rinse handpiece through at least 3 times us-
ing a 20-ml disposable syringe
20 ml
i
Screw scaler instruments onto the rinsing
adapter one after another and rinse through
each of the instruments at least 3 times with
water using a 20-ml disposable pipette
20 ml
Disinfection in an ultrasonic bath
NOTICE
Malfunctions in the handpiece due to
improper handling during cleaning or
disinfecting
i
The handpieces of the Vector must
only be cleaned or disinfected in a
suitable container in an ultrasonic
bath
i
The handpieces must NOT be fully
submerged in fluid
i
Place the removable parts of the handpiece
(Scaler cover, light conductor), tool kit Scaler
(without instruments) in the ultrasonic bath for
EN