Usage
i
Screw scaler instruments onto the rinsing
adapter one after another and rinse through
each of the instruments at least 3 times with
water using a 20-ml disposable pipette
20 ml
Disinfection
i
Place the removable parts of the handpiece
(Scaler cover, light conductor), instrument
holders of the tool kits (without instruments)
and the disassembled handpiece in the clean-
ing bath for the required reaction time, so that
all parts are covered
i
Clean all accessible external surfaces and in-
ternal surfaces below the surface for 5 min-
utes with a sterile cleaning brush until all visi-
ble contaminants have been removed
i
Place instruments in small parts baskets into
the cleaning bath
i
Rinse handpiece through at least 3 times
using a 20-ml disposable syringe
20 ml
i
Place instruments in small parts baskets into
the cleaning bath
i
Rinse handpiece through at least 3 times
using a 20-ml disposable syringe
20 ml
i
Screw Scaler instruments onto the rinsing
adapter one after another and rinse through
each of the instruments at least 3 times using
a 20-ml disposable syringe
20 ml
i
Remove all rinse adapters
i
Comply with the cleaning agents' action times
specified by the manufacturer
Intermediate rinsing
After the action time prescribed by the manufac-
turer:
i
Rinse all components under running water for
at least 1 minute (temperature < 20 °C)
i
Rinse handpiece through at least 3 times us-
ing a 20-ml disposable syringe
20 ml
EN