9000-615-32/02 1808V002
9
Product description
4 Technical data
Electrical data – power supply unit
Rated voltage
V AC
100 - 240
Mains frequency
Hz
50 - 60
Rated power
W
30
Current consumption
A
1 - 05
Protection class
II
Type of protection
IP 20
Electrical data – basic unit and handpieces
Voltage
V DC
24
Electrical power
ON
W
48
Electrical power
Active
W
96
Electrical power
Standby
W
12
Type of protection
IP 20
General technical data – basic unit and handpieces
Operating frequency:
kHz
approx 27–32
Medical device
Class IIa
Duty cycle
%
100
Fluid container fill capacity
ml
600
Scaler handpiece water consumption
ml/min
approx 30–45
Max surface temperature of instruments
°C
58
Weight
Basic unit
kg
2
Scaler handpiece
g
approx 56
Dimensions (W x H x D)
Basic unit
cm
215 x 252 x 16
Scaler handpiece
cm
Ø 20 x 94
Battery for flexible foot switch
Voltage
V
3
Type
Lithium CR2032
Ambient conditions during transport and storage
Temperature
°C
-15 to +60
Relative humidity
%
Max 95
Ambient conditions during operation
Temperature
°C
+10 to +40
Relative humidity
%
Max 80
Electromagnetic compatibility (EMC)
Interference emission measurements
High-frequency emissions in accordance
with CISPR 11
Group 1
Class B
EN