15 Taking out of use
15.1 Taking the unit out of use
If the unit is not to be used for a prolonged
period of time, we recommend that it is properly
shut down and taken out of use.
To do this, the accumulated condensation water
from the unit must be drained.
❯
Switch on the unit and wait until the cut-off
pressure is reached.
Membrane drying unit
❯
While the compressor unit is running, open the
condensate drain valve on the membrane dry-
ing unit. When no more condensation water
emerges, close the condensate drain valve.
❯
Switch off the device.
Pressure tank
❯
Open the condensate drain valve.
Once the start-up pressure has been reached
the compressor will switch on.
❯
With the compressor switched on and the con-
densate drain valve open, wait until no more
condensation water emerges.
❯
Switch off the unit.
❯
Close the condensate drain valve when no
more air escapes.
❯
Disconnect all power from the device.
❯
Disconnect the compressed air connection on
the quick release coupling.
15.2 Storage of the unit
WARNING
Risk of explosion of the pressure tank
and pressure hoses
❯
The pressure tank and the pressure
hoses must be vented before they are
stored or transported.
❯
Protect the unit against moisture, dirt and
extreme temperatures during transport (refer to
the section on "Ambient conditions").
❯
Only store the unit when it has been completely
emptied.
Usage
38