Place a collector tray under the condensate
separator or the membrane drying unit
(depending on type).
9.5
Electrical connections
Safety when making electrical connections
The unit has no main power switch. For
this reason it is important that the unit is
be set up in such a way that the plug can
be easily accessed and unplugged if
required.
❯
The unit must only be connected to a correctly
installed power outlet.
❯
Make sure that none of the electrical cables
leading to the unit are under any mechanical
tension.
❯
Before initial start-up check that the mains sup-
ply voltage and the voltage stated on the type
plate match (see also "4. Technical data").
Establishing the electrical connections
DANGER
Risk of electric shock due to defective
mains cable
❯
Mains cables must not be allowed to
come into contact with any hot surfa-
ces on the unit.
❯
Connect the mains plug to an earthed power
outlet.
10 Commissioning
In many countries technical medical prod-
ucts and electrical devices are subject to
regular checks at set intervals. The owner
must be instructed accordingly.
❯
Turn on the unit power switch or the main sur-
gery switch.
❯
Carry out an electrical safety check in accord-
ance with applicable local regulations (e.g. the
German Ordinance on the Installation, Opera-
tion and Use of Medical Devices / Medizinpro-
dukte-Betreiberverordnung) and record the
results as appropriate, e.g. in the technical log
book.
❯
Carry out and document the instruction and
handover for the unit.
A sample handover report is included in
the attachment.
10.1 Check the motor protection
switch
After installation of the compressor the motor
protection switch needs to be checked and
adjusted if it is not set up correctly. It was set to
the recommended setting in the factory (see "4
Technical data").
❯
Switch on the unit at the pressure switch by
rotating the switch to the "I" position.
Assembly
5152100003L02 1910V001
27
EN