4
Technical data
4.1
Primo
Electrical data
5152-01
Nominal voltage
V
230
Mains frequency
Hz
50
60
Nominal current at 8 bar (0.8 MPa)
A
4.4
4.3
Motor protection switch, recommended
setting
A
5
4.5
Type of protection
IP 24
Mains fuse*
A
10
*
Circuit breaker fuse characteristics B, C or D in accordance with EN 60898-1
General technical data
Pressure tank volume
l
20
Displacement, approx.
l/min
105
125
Delivery at 5 bar (0.5 MPa)*
l/min
58
66
Pressure build-up phase 0-7.5 bar (0-0.75
MPa), approx.
s
160
135
Duty cycle
%
100
Start-up pressure
bar (MPa)
6 (0.6)
Cut off pressure
bar (MPa)
7.8 (0.78)
Cut off pressure, max. adjustable
bar (MPa)
9.5 (0.95)
Safety valve, maximum permissible oper-
ating pressure
bar (MPa)
10 (1)
Pressure dew point at 7 bar (0.7 MPa)**
°C
£
+5
Dimensions (H x W x D)***
cm
69 x 49 x 47
Weight
kg
45
Noise level****
without sound insulation
with sound insulation
dB(A)
dB(A)
65
49
66
50
*
Delivery without membrane drying unit, at +20°C and 1013 mbar (0.1 MPa)
**
Value determined at an ambient temperature of +40 °C
***
Values without accessories and add-on parts
****
Noise level in accordance with ISO 3744
Filter mesh size
Compressor air intake filter
µm
3
Fine filter for membrane drying unit
µm
3
Sterile filter for membrane drying unit
µm
0.01
Sintered filter for membrane drying unit
µm
35
Product description
8
5152100003L02 1910V001
EN