4.3
Duo
Electrical data
5252100001
Nominal voltage
V
110-115
110-127
Mains frequency
Hz
50
60
Nominal current at 8 bar (0.8 MPa)
A
14.1-14.6
14.2-13.3
Motor protection switch, recommended
setting
A
17
16
Type of protection
IP 24
Mains fuse*
A
20
*
Circuit breaker fuse characteristics B, C or D in accordance with EN 60898-1
General technical data
Pressure tank volume
l
20
Suction power, approx.
l/min
210
255
Delivery at 5 bar (0.5 MPa) *
l/min
115
130
Pressure build-up phase 0 - 7.5 bar (0 -
0.75 MPa) approx.
s
80
70
Duty cycle
%
100
Start-up pressure
bar (MPa)
6 (0.6)
Cut-off pressure
bar (MPa)
7.8 (0.78)
Cut-off pressure, max. adjustable
bar (MPa)
9.5 (0.95)
Safety valve, maximum permissible oper-
ating pressure
bar (MPa)
10 (1)
Pressure dew point at 7 bar (0.7 MPa) **
°C
£
+5
Dimensions (H x W x D) ***
cm
69 x 49 x 47
Weight
kg
50
Noise level ****
without sound insulation
with sound insulation
dB(A)
dB(A)
66
55
68
58
*
Delivery without membrane drying unit, at +20°C and 1013 mbar (0.1 MPa)
**
Value determined at an ambient temperature of +40 °C
***
Values without accessories and add-on parts
****
Noise level in accordance with ISO 3744
Filter mesh size
Compressor air intake filter
µm
3
Fine filter for membrane drying unit
µm
3
Sterile filter for membrane drying unit
µm
0.01
Sintered filter for membrane drying unit
µm
35
Ambient conditions during storage and transport
Temperature
°C
-10 to +55
Product description
12
5152100003L02 1910V001
EN