![Durr Dental Duo Installation And Operating Instructions Manual Download Page 31](http://html.mh-extra.com/html/durr-dental/duo/duo_installation-and-operating-instructions-manual_2542878031.webp)
10.4 Draining the condensation
water
Temperature changes during transport may
cause condensation water to accumulate in the
pressure tank. The condensation water can only
be drained from the pressurised pressure tank.
❯
Switch on the unit at the pressure switch and
wait until the cut-off pressure is reached.
❯
At maximum tank pressure, slowly open the
condensate drain valve.
❯
Close the condensate drain valve as soon as
all of the accumulated condensation water has
been blown out.
10.5 Adjusting the rate of flow at
the pressure reducer
The pressure reducer regulates the rate of flow in
the system and adjusts it to the required operat-
ing pressure. In order to adjust the rate of flow air
needs to be extracted via a consumer.
❯
Activate the air consumer unit.
❯
Lift the rotary knob at the pressure reducer.
❯
Adjust the rate of flow via the rotary knob.
Turn the knob in the "+" direction to increase
the rate of flow.
Turn the knob in the "-" direction to reduce the
rate of flow.
❯
Press in the rotary knob until it engages and
cannot be adjusted.
+
-
1.
2.
3.
Assembly
5152100003L02 1910V001
29
EN