4.4
Trio
Electrical data
5352-01
5352100001
Nominal voltage
V
230
Mains frequency
Hz
50
Nominal current at 8 bar (0.8 MPa)
A
8.6
Motor protection switch, recommended
setting
A
8.6
Type of protection
IP 24
Mains fuse*
A
10
Max. permissible mains impedance in
accordance with EN 61000
‑
3
‑
11
W
£
0.078
*
Circuit breaker fuse characteristics B, C or D in accordance with EN 60898-1
General technical data
Pressure tank volume
l
50
Suction power, approx.
l
315
Delivery at 5 bar (0.5 MPa) *
l/min
160
Pressure build-up phase 0 - 7.5 bar (0 -
0.75 MPa) c.
s
140
Duty cycle
%
100
Start-up pressure
bar (MPa)
6 (0.6)
Cut-off pressure
bar (MPa)
7.8 (0.78)
Cut-off pressure, max. adjustable
bar (MPa)
9.5 (0.95)
Safety valve, maximum permissible oper-
ating pressure
bar (MPa)
10 (1)
Pressure dew point at 7 bar (0.7 MPa) **
°C
£
+5
Dimensions (H x W x D) ***
cm
76 x 74 x 52
Weight
kg
80
Noise level ****
without sound insulation
with sound insulation
dB(A)
dB(A)
67
54
*
Delivery without membrane drying unit, at +20°C and 1013 mbar (0.1 MPa)
**
Value determined at an ambient temperature of +40 °C
***
Values without accessories and add-on parts
****
Noise level in accordance with ISO 3744
Filter mesh size
Compressor air intake filter
µm
3
Fine filter for membrane drying unit
µm
3
Sterile filter for membrane drying unit
µm
0.01
Product description
14
5152100003L02 1910V001
EN