5 – Performance and Limitations
Operating Ranges
Hematology parameters measured with the QBC Star system are considered valid over the following
range of values:
Hematocrit
15
–
65%
Hemoglobin
5.0
–
20.0
g/dL
Mean Corpuscular Hemoglobin Concentration
25 - 37.3 g/d
Platelet Count
20 – 999 x 10
9
/L
White Blood Cell Count
1.6 – 99.9 x 10
9
/L
Granulocyte Count
0.8 – 70.0 x 10
9
/L
Granulocyte
%
1
-
99%
Lymph/Mono Count
0.8 – 99.9 x 10
9
/L
Lymph/Mono
% 1-99%
Results that fall outside these ranges may be confirmed by alternate methods.
6000-300-000 5-1
As a reference, the STAR will display the
numerical values of some parameters when
they are slightly outside the Operating Ranges.
On the printout, these values will be highlighted
with double asterisks and marked as “**
Reference Result” below the area of results. On
the LCD display, these values will be flashing.
If the results are much farther outside
the Operating Ranges, the values will be
withheld and the phrase “Out of Operating
Range” will be printed on the printout
instead.
Summary of Contents for QBC STAR
Page 14: ...THIS PAGE INTENTIONALLY LEFT BLANK...
Page 28: ...THIS PAGE INTENTIONALLY LEFT BLANK...
Page 40: ...QBC STAR System Operator s Manual 6000 300 000 4 12 THIS PAGE INTENTIONALLY LEFT BLANK...
Page 50: ...THIS PAGE INTENTIONALLY LEFT BLANK...
Page 72: ...6000 300 000 A 1 8 Appendices 6000 300 000 8 1...
Page 74: ...THIS PAGE INTENTIONALLY LEFT BLANK...
Page 76: ...THIS PAGE INTENTIONALLY LEFT BLANK...
Page 78: ...THIS PAGE INTENTIONALLY LEFT BLANK...
Page 80: ...THIS PAGE INTENTIONALLY LEFT BLANK...