Copyright 2010 Baker Hughes Company.
50 | PACE Indicators Instruction Manual–English
Chapter
6.
Reference
6.6.3 Gas Head Correction
Corrects pressure reading for the height difference between instrument reference level and UUT.
For accuracy, head correction must be enabled and the parameters set for each sensor.
•
For UUT positioned higher than the reference level of the PACE, enter a positive height
correction.
•
For UUT positioned lower than the reference level of the PACE, enter a negative height
correction.
•
When calibrating the PACE instrument disable the gas head correction and correct the
actual applied pressures for height.
6.6.4 Lock Tasks
6.6.4.1 Individual Tasks
Allows any combination of individual tasks to be disabled.
Note:
Restricts operation of the instrument to specific tasks or functions, recommended for
production procedures.
6.6.4.2 All
Disables all tasks.
6.6.5 Change PIN
Changes the Supervisor PIN: enter the existing PIN, then the new PIN and confirmation of the
new PIN.
Note:
Confirmation of the new PIN permanently replaces the old PIN. Record this new PIN and
keep in a safe place. If new PIN is lost it can only be reset by returning the instrument to a Druck
service center.
6.6.6 User Defined Units
Permits the user to define a set of units. Following the on-screen prompts special units may be
set by selecting a Pascal multiplier and assigning a five character name.
6.6.7 Instrument Alias Name
Permits the user to define a 20 character alias name for the instrument. The instrument returns
this name through the communications interfaces.
6.6.8 Language
Operation in any of the following languages can be selected:
•
English (default)
•
French
•
German
•
Italian
1
Reference level.
1
Summary of Contents for PACE1000
Page 1: ...Druck com PACE Indicators Pressure Automated Calibration Equipment Instruction Manual...
Page 2: ......
Page 10: ...Copyright 2010 Baker Hughes Company viii PACE Indicators Instruction Manual English...
Page 29: ...Copyright 2010 Baker Hughes Company English PACE Indicators Instruction Manual 19 Data Logging...
Page 75: ......