48 | PACE Indicators Instruction Manual–English
Chapter
6.
Reference
d. Use the number touch pad at the bottom of the screen to input the new IP address and
press the CONTROLLER IP ADDRESS touch pad at the top of the screen to set the
new IP address.
11. To view the recent IP addresses, complete the following:
a. On the ETHERNET PARAMETER screen, use the UP and DOWN arrows on the right
of the screen to highlight the ACCESS CONTROL field.
b. Press the ACCESS CONTROL button on the top of the screen.
c.
Use the UP and DOWN arrows to highlight the SELECT FROM RECENT field.
d. Press the SELECT FROM RECENT button at the top of the screen to view the IP
address history.
6.6.2.6 Comms Range Setup
A Comms Range value assigns a unique index number to a pressure sensor installed in the
PACE. This is used to identify the pressure sensor when communicating with the PACE using
SCPI. For example, a PACE is installed with three pressure sensors:
•
1150
mbar
a
•
2
bar
g
•
Barometric sensor