Copyright 2010 Baker Hughes Company.
English–PACE Indicators Instruction Manual | 49
Supervisor Setup
In the Comms Range value configuration, they are assigned the following ranges:
•
Range 1 = 1150
mbar
a
•
Range 2 = Barometric sensor
•
Range 3 = 2
bar
g
When communicating with the PACE, to request a pressure reading from the 1150
mbar
a
sensor, the SCPI command requests a reading from Index 1. To request a pressure reading from
the Barometric sensor, the SCPI command requests a reading from Index 2. Etc.
Note:
The following procedure only allows the supervisor to select these user pre-installed
values. To initially define, change or delete the Comms Range values, refer to document K0472
PACE SCPI Remote Communications Manual.
To configure the Comms Range Setup:
1. Navigate to the supervisor communications configuration. Refer to Section
“Navigating to Communications Menu,” on page
2. On the COMMUNICATIONS screen, select COMMS RANGE SETUP.
3. On the COMMS RANGE SETUP, use the Up and Down arrows to highlight the desired
Range (Ranges 1 to 10).
Comms Range Setup is used to align the SCPI index (x) to a pressure range.
Where Range is index (1)… etc.
4. Press the top touch area on the screen to change the range value.
5. Use the Up and down arrow to highlight the new Range value.
6. Press the top Measure on the screen to change the range value. The new range value is set
and the screen returns to the COMMS RANGE SETUP screen.
7. If necessary, repeat Steps 3 to 6 to set up other range values.
8. Press the Escape icon to go back to the COMMS RANGE SETUP screen.
9. When complete, press the Escape icon as necessary to return to the Measure screen.
Summary of Contents for PACE1000
Page 1: ...Druck com PACE Indicators Pressure Automated Calibration Equipment Instruction Manual...
Page 2: ......
Page 10: ...Copyright 2010 Baker Hughes Company viii PACE Indicators Instruction Manual English...
Page 29: ...Copyright 2010 Baker Hughes Company English PACE Indicators Instruction Manual 19 Data Logging...
Page 75: ......