Copyright 2010 Baker Hughes Company.
40 | PACE Indicators Instruction Manual–English
Chapter
6.
Reference
6.6.2.2 IEEE
488
Located on the rear panel an external IEEE
488 connection has the following configuration:
To configure the IEEE
488 connection:
1. Navigate to the supervisor communications configuration. Refer to Section
“Navigating to Communications Menu,” on page
2. On the COMMUNICATIONS screen, select IEEE488.
3. On the IEEE488 PARAMETERS screen, use the Up and Down arrows to highlight and edit
the desired parameter.
6.6.2.3 RS
-
232
Located on the rear panel an external RS
-
232 connection has the following configuration:
Table
6
-
1: IEEE
488 Configuration Options
Item
Description
Connector
24-way ‘D’ female wired as IEEE
488 standard.
Communications
IEEE
488 GPIB
Default Address
16
Protocols
SCPI
Heritage Emulation
DPI 142/150, DPI 141
Table
6
-
2: RS
-
232 Configuration Options
Item
Description
Connector
9-way ‘D’ female. For pin connections, refer to Table
11.
Communications
RS
-
232 point-to-point only. Daisy chain is not supported.
Baud Rate Power-up Default
9600, no parity, handshake = Xon/Xoff
Baud Rates Selectable
a
a. Selectable through the user interface.
2400, 4800, 9600, 19k2, 38k4, 57k6, 115k2
Parity
None, Odd, Even
Flow Control
None, Hardware, Xon/Xoff
Protocols
SCPI
Heritage Emulation
DPI 142/150, DPI 141
Terminator
CR or LF or CR/LF
Summary of Contents for PACE1000
Page 1: ...Druck com PACE Indicators Pressure Automated Calibration Equipment Instruction Manual...
Page 2: ......
Page 10: ...Copyright 2010 Baker Hughes Company viii PACE Indicators Instruction Manual English...
Page 29: ...Copyright 2010 Baker Hughes Company English PACE Indicators Instruction Manual 19 Data Logging...
Page 75: ......