background image

24

Dräger Polytron 8000

Settings menu

Testing the maintenance signal

This function is used to set the analog interface for the

maintenance signal.
1. Select 

Settings > Communication > Analog interface >

Set mainten. 

and confirm. 

2. Select 

On

 or 

Off

 and confirm.

The current for the maintenance signal will be transmitted

on the analog interface. 

8.3.3

Profibus address

Not yet active - for future functions.

8.4

Sensor settings

8.4.1

Sensor change

A sensor can be changed during operation with this function

without triggering a fault signal at the central controller. It also

ensures that all sensor data in the microprocessor can be

saved in sensor memory (EEPROM) before disconnecting

the sensor.

A sensor can be changed at any time. However, for safety

reasons a fault alarm is generated when the sensor is

disconnected until a new sensor is connected.
1. Select 

Settings > Sensor > Sensor change 

and confirm.

The maintenance signal is generated at the analog

interface, the display shows:  "Please remove the sensor".

2. Replace old sensor with a new sensor, see chapter 8.4.1

on page 24.

When the new sensor is installed:

Display: "Loading data, please wait".

When the sensor data are loaded:

Display: "Data loaded".

3. Select 

Back to menu

 and confirm.

The maintenance signal at the analog interface remains

pending until the sensor is warmed-up.

The warm-up time depends on the type of sensor and its

history. For example, if the sensor was already warmed-up at

another transmitter and operation only interrupted for a short

time, the warm-up time can be shorter than the time specified

in the Instructions for Use for the sensor. 

Maximum warm-up time of a new sensor:  see the Instructions

for Use for the sensor. 

If an identical sensor (with the same part number) was installed

immediately before, the transmitter configuration is retained

(gas type, measurement range, test gas, calibration interval).

Otherwise the preset values of the sensor (see Instructions for

Use for the sensor) are uploaded to the transmitter if the sensor

lock function (see chapter 8.4.4 on page 24) is deactivated.

8.4.2

Switching Auto CAL on or off 

1. Select 

Settings > Sensor > Set Auto CAL

 and confirm.

2. Select 

On

 or 

Off

 and confirm.

8.4.3

Sensor test (only with sensor test dongle)

Switching automatic sensor test on or off

When the sensor test is enabled, the transmitter continuously 

checks the sensor to ensure reliable functioning. If the sensor 

does not pass the test, an appropriate warning or fault indication 

is issued.
1. Select 

Settings> Sensor > Sensor test 

and confirm.

2. Select 

On

 or 

Off

 and confirm.

Manually triggering the sensor test

The sensor test is started manually with this function:
1. Select 

Settings > Sensor > Set sensor test > 

and confirm.

If a sensor test can be triggered depending on the sensor or

transmitter status: 
2. Select 

Start sensor test 

and confirm.

The result of the sensor test is displayed after a few seconds.

8.4.4

Sensor lock

If the sensor lock is to be switched on or off for the sensor:
1. Select 

Settings > Sensor > Sensor lock 

and confirm.

2. Select 

Activate

 or 

Deactivate

 and confirm.

8.4.5

Changing the gas settings

Settings for gas type, measuring range and measurement

units can be made in this menu.
1. Select 

Settings > Sensor > Gas setting 

and confirm.

2. Select the target gas from the list and confirm.

The measurement units display will be shown.

3. Select the measurement units and confirm.

The full scale reading display will be shown.

4. Set the full scale reading and confirm.

The setting for the full scale reading is displayed. 

5. Select 

Next

 and confirm.

an overview of the new gas settings is displayed.

z

Return to the previous view with 

Previous

 or confirm the

settings with 

Confirm

.

NOTICE

The auto-calibration from the 

Calibration

 menu is only

available if this function is enabled.

Activate

= The transmitter will only accept a new sensor 

if the Part Number (= Dräger Order Number), 

and hence the sensor type, is identical to that 

of the previous sensor.

Deactivate

= The transmitter will also accept other 

sensor types and with thus upload the 

default settings for the sensor, i.e., in this 

case the configuration of the transmitter 

will be altered.

i

i

Summary of Contents for Polytron 8000

Page 1: ...Instructions for Use The user must fully understand and strictly observe the instructions Use the product only for the purposes specified in the Intended use section of this document Dr ger Polytron...

Page 2: ......

Page 3: ...ngles 11 4 Operation 12 4 1 Operating fundamentals 12 4 2 Display and LEDs 12 4 3 Activating the Info mode 13 4 4 Switching to the Quick Menu mode 14 4 5 Switching to Menu mode password protected 14 4...

Page 4: ...ransmitter for explosion protection according to Annex II Sections 1 5 5 1 5 6 and 1 5 7 of Directive 94 9 EC is not covered at present z Exchanging components may compromise the intrinsic safety This...

Page 5: ...ronics and optional relay 4 Lower part of housing 5 Sensor 6 Assembled device 2 1 2 Explosion proof device with e Box for increased safety 1 Housing cover 2 Handle 3 Enclosure with main electronics an...

Page 6: ...which gases of Explosion Groups IIA IIB or IIC and Temperature Classes T4 or T6 depending on maximum ambient temperature or dust of Groups IIIA IIIB or IIIC may occur The temperature range specified...

Page 7: ...with protective ground or chapter 3 6 on page 9 z The connecting wires for the optional relay module must be selected and fused according to the rated voltages currents and environmental conditions z...

Page 8: ...rt the main electronics back in the housing 10 Screw the cover on again and tighten the set screw 3 5 2 Version with relays When the relay module is fitted the wires for the alarm device are connected...

Page 9: ...set screw 2 2 mm Allen screw of the sensor bayonet ring until the bayonet ring is locked to prevent accidental loosening NOTICE For clear and far ranging detection of a device fault an alarm device s...

Page 10: ...the remote adapter 4 in the opening of the EC sensing head local The Dr ger label on the sensor adapter must point to the mark on the sensing head housing 5 5 Lock sensor adapter with bayonet ring 6 F...

Page 11: ...r the Polytron 8000 To install a SW dongle 1 Switch off the power to the unit or declassify the area according to the local regulations 2 Release the set screw and unscrew the housing cover from the d...

Page 12: ...4 2 2 on page 13 z SIL when SIL status is activated When the optional relay PCB is used z If the first alarm has been triggered the red LED flashes single and A1 is displayed z If the second alarm has...

Page 13: ...terface Only displayed when the 4 20 mA interface is enabled z A1 Alarm limit and measurement units1 z A2 Alarm limit and measurement units1 Warning message present To display warnings see chapter 7 1...

Page 14: ...D test software version and initialization and starts the warm up phase The display indicates that the sensor will be ready to operate in hh mm ss the countdown begins and the Polytron 8000 sends the...

Page 15: ...ng safety data sheets Provide venting into a fume hood or outside the building Always calibrate the zero point first before the sensitivity Otherwise the calibration will be faulty NOTICE Calibration...

Page 16: ...s Sensor Autocal set to enable or disable the function 2 Supply fresh air or apply nitrogen or compressed air to the sensor at a flow rate of 0 5 L min 3 Select Calibration Auto cal and confirm Please...

Page 17: ...age 20 Datalogr on off SW dongle page 21 Set alarm A2 page 20 Set Datalogger Display page 21 Set fault page 20 Clear datalogr Calibration PWD page 20 Sample time Settings PWD page 20 Peak Average Trig...

Page 18: ...form If all the numerical groups are displayed as 00 there are no faults or warnings present 7 1 4 Displaying installed modules Display of the installed hardware modules z Select Information Instrume...

Page 19: ...threshold and confirm The setting for the alarm threshold is displayed 4 Select Next and confirm The setting for the alarm direction is displayed 5 Select Rising or Falling and confirm 6 Select Next a...

Page 20: ...nfiguring the function key This function is used to assign a predefined function to the key The selected function is activated by briefly touching the key 1 Select Settings Instrument Function key and...

Page 21: ...tings Instrument Display Displaytest and confirm 2 Select On or Off and confirm When this function is activated the display is shown inverted and all the LEDs are lit 8 3 Communication settings 8 3 1...

Page 22: ...confirm 3 Set the current and confirm 4 Select Confirm and confirm Switching the warning signal on or off This function is used to switch the warning signal on the analog interface on and off If an e...

Page 23: ...to set various concentrations in the range between 0 and 100 of the selected unit of measure The current output is set to correspond to the selected concentration 1 Select Settings Communication Anal...

Page 24: ...aded to the transmitter if the sensor lock function see chapter 8 4 4 on page 24 is deactivated 8 4 2 Switching Auto CAL on or off 1 Select Settings Sensor Set Auto CAL and confirm 2 Select On or Off...

Page 25: ...mA interface 8 4 6 Resetting the sensor to factory settings This function resets all the sensor parameters to the factory default settings see chapter 12 3 on page 32 1 Select Settings Sensor Sensor...

Page 26: ...he trigger threshold The trigger threshold is specified as a percentage of the maximum range Example A trigger threshold of 2 is entered for a maximum range of 500 ppm Consequently the datalogger will...

Page 27: ...nterface and reconnect the power supply For operation without controller connect terminal 3 with terminal 4 see chapter 3 5 1 on page 8 045 Device not detecting any sensor Check contacts If this fault...

Page 28: ...e device to factory settings see chapter 8 2 6 on page 21 If this fault occurs again Have the device checked by Dr gerService 104 No valid date or time set Set time and date see chapter 8 2 3 on page...

Page 29: ...m packaging 4 Insert sensor or MEC sensor adapter 4 into the opening The Dr ger label on the sensor must point to the mark on the sensing head housing 5 5 Lock sensor with bayonet ring 6 For operation...

Page 30: ...waste disposal regulations For information consult your local environmental agency local government offices or appropriate waste disposal companies CAUTION When the main electronics is changed all se...

Page 31: ...10 conductor resistance Operating current max 80 mA at 24 V DC no relay sensor not remote 100 mA at 24 V DC with relay remote sensor Relay design optional Single pole double throw 5 A at 230 V AC 5 A...

Page 32: ...n AWG Cable size in mm2 Power supply and signal 4 4 7 0 0 5 0 8 Nm 24 12 0 2 2 5 Relay 4 4 7 0 0 5 0 8 Nm 20 12 0 5 2 5 Menu Standard setting A1 Alarm Depending on the sensor A2 Alarm A1 Relay alarm D...

Page 33: ...2 alarm is present If an alarm is configured as acknowledgeable and the alarm is acknowledged the LED lights Alarm hierarchy A2 has a higher priority for the LED than A1 However the A1 and A2 relays o...

Page 34: ...on adapter PE Europe 68 06 978 Calibration adapter Viton 68 10 536 Pipe mounting kit Polytron 5000 8000 45 44 198 Duct mount kit 68 12 725 EC Sensor Head Remote Polytron 8000 with wall pipe mounting k...

Page 35: ...Declaration of conformity Dr ger Polytron 8000 35 14 Declaration of conformity...

Page 36: ...36 Dr ger Polytron 8000 15...

Page 37: ...www norrscope com Dr ger Polytron 8000 37...

Reviews: