10
Dräger Polytron 8000
Installation
3.8
Installation of EC sensing head remote
3.8.1
Wall-pipe mounting set
1. Install wall-pipe mount (8) as specified by the assembly
instructions.
2. Screw the EC sensing head to the wall mount with the
included screw and washer (6).
3. Select the appropriate protection cap (7) and place on
the screw.
3.8.2
Installing sensor
1. Unscrew bayonet ring (3) from EC sensing head and
remove blank.
2. Remove sensor from packaging.
3. Insert sensor or MEC sensor adapter (4) into the opening.
The Dräger label on the sensor must point to the mark on
the sensing head housing (5).
4. Lock sensor with bayonet ring.
5. Screw in the set screw (2; 2 mm Allen screw) of the sensor
bayonet ring until the bayonet ring is locked to prevent
accidental loosening.
3.9
Connecting to the Polytron 8000
1. Attach remote cable plug (11, order no. see chapter 13 on
page 34) to the EC sensing head remote (12) and lock by
rotating the ring clockwise.
2. Screw the cable terminal of the shield (13) to the hole
(M5 thread) on the wall mount (14) for the ground.
3. Unscrew bayonet ring (3) from EC sensing head local and
remove blank. The black bayonet ring can be replaced with
the silver bayonet ring for an improved appearance.
4. Insert the remote adapter (4) in the opening of the EC
sensing head local. The Dräger label on the sensor adapter
must point to the mark on the sensing head housing (5).
5. Lock sensor adapter with bayonet ring.
6. Fasten the cable terminal of the remote cable to the device
housing for a ground (15).
7. Screw in the set screw (2; 2 mm Allen screw) of the sensor
bayonet ring until the bayonet ring is locked to prevent
accidental loosening.
CAUTION
Note the full range of environmental influences to which
the sensing head may be exposed. Avoid external
influences such as splashwater, oil, corrosive aerosols
(salt spray) etc. and the possibility of mechanical
damage.
The gas inlet area of the sensor must be protected
from water, dust and mechanical damage and kept
clear of dirt. Do not allow paint to block the gas inlet if
painting is conducted in the area.
Multiple EC sensing heads remote must not be
connected in series.
NOTICE
The EC measuring sensor wall and pipe mounting
accessory set (order no. 45 44 213) is required to
mount the EC measuring sensor on a wall or pipe.
The sensing head should be mounted at a low-
vibration location at even a temperature near
a potential leakage point (avoid direct sunlight).
NOTICE
Observe the direction of the remote connector!
!
i
i
00233298.eps
8
9
8
9
8
3
4
5
6
7
2
14
i
i
NOTICE
The EC sensing head is automatically recognized by
the transmitter.
00333298.eps
2
2
3
5
11
14
13
12
4
10
15
i
i