26
Dräger Polytron 8000
Settings menu
8.5
Datalogger
The datalogger has a storage capacity of approx.
3000 measurements. At a sampling interval of 1 minute,
this would allow monitoring for approx. 50 hours. The monitoring
time can be significantly increased if the
Trigger
function
is enabled.
The results in the datalogger can only be evaluated with the
PolySoft 8000 PC software. An evaluation of the datalogger
contents over the last 15 minutes is available in the menu
under
Information > Datalogger > Graph
(see chapter 7.3.2
on page 18).
8.5.1
Switching the datalogger on or off
1. Select
Settings > Datalogger > Datalogr. on/off
and
confirm.
2. Select
On
or
Off
and confirm.
8.5.2
Setting up the datalogger
Setting the sampling interval
1. Select
Settings > Datalogger > Datalogr. on/off >
Sample time
and confirm.
2. Select the sample time and confirm.
Setting peak/average
1. Select
Settings > Datalogger > Set datalogger > Peak/
average
and confirm.
2. Select the desired function and confirm.
Switching the trigger on or off
1. Select
Settings > Datalogger > Set datalogger > Trigger
on/off
and confirm.
2. Select
On
or
Off
and confirm.
Setting the trigger threshold
The trigger threshold is specified as a percentage of the
maximum range.
Example: A trigger threshold of 2 % is entered for a maximum
range of 500 ppm. Consequently the datalogger will only store
measurements if they deviate (relative to the last stored value)
by 10 ppm.
1. Select
Settings > Datalogger > Datalogr. on/off >
Trigger value
and confirm.
2. Set the trigger value and confirm.
Setting the storage mode
1. Select
Settings > Datalogger > Set Datalogger > Stack/
Roll
and confirm.
2. Select the desired condition and confirm
8.5.3
Clearing the datalogger
1. Select
Settings > Datalogger > Clear Datalogr.
and confirm.
2. To clear the datalogger, select
Confirm
and confirm.
Peak value
The measured maximum value (for O
2
, the
minimum value) of the concentration within
the selected sampling interval is stored.
Average
The average value of all the concentrations
measured within the selected sampling time
is stored.
On
Concentrations will only be stored if they
exceed (relative to the last stored value)
a threshold set under Trigger Value.
Off
The measurements made within the set
sampling time are always stored.
Overwrite
When the capacity of the datalogger is
used up, the older data will be overwritten.
Hold
When the capacity of the datalogger is
used up, no more data can be stored.
An appropriate warning is issued.