Cleaning, Disinfection and Sterilization
96
Instructions for Use Oxylog 3000 SW 1.n
Disassembly
Disassemble the reusable hose system
1
Disconnect the ventilation hose (A) from the gas
output.
2
Disconnect the flow measuring hoses (B) from
the nozzles.
3
Disconnect the medical gas hose (C) from the
Oxylog 3000
.
046
B
A
C
048
CAUTION
When disconnecting the ventilation hose, always
grip the sleeve (A) and not the corrugations (B)!
If this is not done, the corrugations or hose may
be torn from the sleeve.
A
B