Instructions for Use Oxylog 3000 SW 1.n
117
Technical Data
Operating data
Power supply
Power supply
Input voltage
Oxylog 3000
19 V ±0.5 V DC
With DC/DC converter
12 / 24 / 28 V DC
Current consumption
With battery charge
max. 3.8 A, typically 2.1 A
Operating time with fully
charged internal battery
without mains supply for
"typical" ventilation
approximately 4 hours
Battery type
Lithium ion battery
Charging times
The specified charging time applies when recharging
the battery completely after it has been depleted.
Lithium ion battery
Approx. 5 hours
Permissible ambient temperature during charging
0
o
C to 35
o
C
Indication of battery capacity
in 25% increments.
Accuracy of the capacity indication
The indicated capacity is determined by the battery
itself. The accuracy depends on the type and manu-
facturer and may deteriorate with frequent partial
discharge and during operation in extreme tempera-
tures. The internal battery is only reconditioned after
being discharged completely and recharged at room
temperature 25
o
C.
The criteria for the warnings
!!! Int. battery dis-
charged
and
!! Charge int. battery
are therefore
based on measurement of the battery voltage. The
capacity indicated at this moment may differ from the
actual capacity of the internal battery.
Battery storage time
The internal battery must always be removed from
the ventilator for storage and recharged completely
after 12 months at the latest (e.g. in the external
battery charging station).
Summary of Contents for Oxylog 3000
Page 26: ...26 Instructions for Use Oxylog 3000 SW 1 n This page intentionally left blank ...
Page 44: ...44 Instructions for Use Oxylog 3000 SW 1 n This page intentionally left blank ...
Page 74: ...74 Instructions for Use Oxylog 3000 SW 1 n This page intentionally left blank ...
Page 86: ...86 Instructions for Use Oxylog 3000 SW 1 n This page intentionally left blank ...
Page 102: ...102 Instructions for Use Oxylog 3000 SW 1 n This page intentionally left blank ...
Page 106: ...106 Instructions for Use Oxylog 3000 SW 1 n This page intentionally left blank ...
Page 110: ...110 Instructions for Use Oxylog 3000 SW 1 n This page intentionally left blank ...
Page 134: ...134 Instructions for Use Oxylog 3000 SW 1 n This page intentionally left blank ...
Page 137: ...Instructions for Use Oxylog 3000 SW 1 n 137 This page intentionally left blank ...