Instructions for Use Oxylog 3000 SW 1.n
93
Problem Solving
Messages in the information window
(Numerical examples)
Message
Cause
Explanation/Remedy
f = 12 per min or
VT = 800 mL
I:E = 1 : 1.5
Flow = 15 L/min
Change in Tinsp, f or VT in ventila-
tion mode SIMV.
f = 12 per min or
VT = 800 mL
Tinsp= 0.7 s
Flow = 35 L/min
Change in I:E, f or VT in ventilation
mode IPPV (CMV), IPPVAssist
(CMVAssist).
I:E = 1 : 1.5
Texp= 2 s
Change in Tinsp or f in ventilation
mode BIPAP
®
(PCV+).
PEEP >10 mbar has been set but
not confirmed.
The required setting of
PEEP >10 mbar is only possible
when confirmed via the rotary
knob.
Gas consumption = 10 L/min
Standard display in information
window for the current gas
consumption.
(Battery capacity)
Standard display in information
window for the current battery
capacity.
Pinsp>=PEEP + 3 mbar
Set PEEP + 3 mbar > Pinsp
Set Pinsp > PEEP + 3 mbar
PASB = 22 mbar
Change in
Δ
ASB or PEEP.
PASB
is the absolute pressure
resulting from PEEP +
Δ
ASB.
Confirm PEEP above 10 mbar ?
Summary of Contents for Oxylog 3000
Page 26: ...26 Instructions for Use Oxylog 3000 SW 1 n This page intentionally left blank ...
Page 44: ...44 Instructions for Use Oxylog 3000 SW 1 n This page intentionally left blank ...
Page 74: ...74 Instructions for Use Oxylog 3000 SW 1 n This page intentionally left blank ...
Page 86: ...86 Instructions for Use Oxylog 3000 SW 1 n This page intentionally left blank ...
Page 102: ...102 Instructions for Use Oxylog 3000 SW 1 n This page intentionally left blank ...
Page 106: ...106 Instructions for Use Oxylog 3000 SW 1 n This page intentionally left blank ...
Page 110: ...110 Instructions for Use Oxylog 3000 SW 1 n This page intentionally left blank ...
Page 134: ...134 Instructions for Use Oxylog 3000 SW 1 n This page intentionally left blank ...
Page 137: ...Instructions for Use Oxylog 3000 SW 1 n 137 This page intentionally left blank ...