Alarms
Setting alarm limits
Upper alarm limit for Paw
Pressure limitation with Pmax
Regardless of the set ventilation mode, the airway
pressure is controlled by the ventilator and limited
to the set maximum inspiratory pressure Pmax.
Pmax appears in the pressure curve as a dashed
line.
When this dashed line is reached, Oxylog 3000
issues a
!!! Paw high
alarm. The volume-controlled
stroke is terminated (ventilation with nonconstant
volume).
1
Set the maximum airway pressure Pmax via the
Pmax
control (A).
The airway pressure is limited when Pmax is rea-
ched; inspiration will not be terminated prematurely.
Lower alarm limit for Paw
A lower alarm limit need not be set for the airway
pressure Paw. Oxylog 3000 automatically genera-
tes an alarm when it no longer detects a pressure
difference of more than 5 mbar between inspiratory
and expiratory pressure.
To set alarm limits for MV and fspn
Display (example):
Alarms
Alarms
screen with variable alarm limits.
= lower alarm limit.
= upper alarm limit.
2
Press the key
Alarms
(B).
Example: Setting the upper alarm limit for MV.
1
Select and activate the line
MV
on the
display.
2
Set and confirm the value.
CAUTION
Set alarm values carefully.
Extreme alarm values can render the alarm
system useless.
039
Oxylog 3000
A
B
11
0371
71-2
Alarm
Range
MV
2 to 41 L/min
MV
0.5 to 40 L/min
fspn
10 to 100 /min
WARNING
Set the lower alarm limit MV
according to
the minimum ventilation required for the
patient.
Otherwise, there is a risk of the patient
receiving insufficient ventilation.
Summary of Contents for Oxylog 3000
Page 26: ...26 Instructions for Use Oxylog 3000 SW 1 n This page intentionally left blank ...
Page 44: ...44 Instructions for Use Oxylog 3000 SW 1 n This page intentionally left blank ...
Page 74: ...74 Instructions for Use Oxylog 3000 SW 1 n This page intentionally left blank ...
Page 86: ...86 Instructions for Use Oxylog 3000 SW 1 n This page intentionally left blank ...
Page 102: ...102 Instructions for Use Oxylog 3000 SW 1 n This page intentionally left blank ...
Page 106: ...106 Instructions for Use Oxylog 3000 SW 1 n This page intentionally left blank ...
Page 110: ...110 Instructions for Use Oxylog 3000 SW 1 n This page intentionally left blank ...
Page 134: ...134 Instructions for Use Oxylog 3000 SW 1 n This page intentionally left blank ...
Page 137: ...Instructions for Use Oxylog 3000 SW 1 n 137 This page intentionally left blank ...