Configuration
80
Instructions for Use Oxylog 3000 SW 1.n
To enter customer service mode
1
Turn controls (A) and (B)
VT
and
Freq.
all the
way to the right.
2
Switch ON (C) the device (briefly press
O
key)
and simultaneously press and hold the
Curves
key (D) and the
Values
key
(E)
until the main
Customer Service Mode
menu
appears.
3
Set the number of the required test in the main
menu with the central rotary knob.
4
Activate test = press rotary knob.
Settings in customer service mode
1
Select the required function with the cursor
(asterisk).
– To select the parameter: turn the rotary
knob.
– To activate the parameter: press the rotary
knob.
– To set the value: turn the rotary knob.
– To confirm the value: press the rotary knob.
To exit the parameter settings menu
1
Select the line
EXIT
.
2
Press the rotary knob and confirm.
The set values are saved and remain effective
whenever ventilation is started after switching
ON.
07737170
scr
een 18
Oxylog 3000
A
B
C
D
E
Curves
11
837170
Summary of Contents for Oxylog 3000
Page 26: ...26 Instructions for Use Oxylog 3000 SW 1 n This page intentionally left blank ...
Page 44: ...44 Instructions for Use Oxylog 3000 SW 1 n This page intentionally left blank ...
Page 74: ...74 Instructions for Use Oxylog 3000 SW 1 n This page intentionally left blank ...
Page 86: ...86 Instructions for Use Oxylog 3000 SW 1 n This page intentionally left blank ...
Page 102: ...102 Instructions for Use Oxylog 3000 SW 1 n This page intentionally left blank ...
Page 106: ...106 Instructions for Use Oxylog 3000 SW 1 n This page intentionally left blank ...
Page 110: ...110 Instructions for Use Oxylog 3000 SW 1 n This page intentionally left blank ...
Page 134: ...134 Instructions for Use Oxylog 3000 SW 1 n This page intentionally left blank ...
Page 137: ...Instructions for Use Oxylog 3000 SW 1 n 137 This page intentionally left blank ...