Instructions for Use Oxylog 3000 SW 1.n
35
Assembly
Oxylog 3000 can be supplied with either O
2
or
medical air.
z
Set the gas type in the configuration menu to
ensure correct metering, "Check connections"
on page 42.
Supply from an O
2
cylinder
1
Use a full O
2
cylinder.
2
Connect the pressure reducer (270 to 600 kPa
delivery pressure, 500 kPa nominal pressure)
to the O
2
cylinder.
3
Connect the O
2
medical gas hose (A) to the
Oxylog 3000.
4
Connect the O
2
medical gas hose to the
pressure reducer (B).
WARNING
No smoking or open flames.
O
2
enhances combustion of other substances
and can intensify fires.
WARNING
Only use compressed gas cylinders and pres-
sure reducers, which comply with all applica-
ble regulations and have been approved.
WARNING
Only use a pressure reducer with a relief valve
at the outlet to limit the delivery pressure to a
maximum of 1000 kPa in case of a malfunc-
tion, to prevent damage to the ventilator!
018
O
2
A
B
C
Summary of Contents for Oxylog 3000
Page 26: ...26 Instructions for Use Oxylog 3000 SW 1 n This page intentionally left blank ...
Page 44: ...44 Instructions for Use Oxylog 3000 SW 1 n This page intentionally left blank ...
Page 74: ...74 Instructions for Use Oxylog 3000 SW 1 n This page intentionally left blank ...
Page 86: ...86 Instructions for Use Oxylog 3000 SW 1 n This page intentionally left blank ...
Page 102: ...102 Instructions for Use Oxylog 3000 SW 1 n This page intentionally left blank ...
Page 106: ...106 Instructions for Use Oxylog 3000 SW 1 n This page intentionally left blank ...
Page 110: ...110 Instructions for Use Oxylog 3000 SW 1 n This page intentionally left blank ...
Page 134: ...134 Instructions for Use Oxylog 3000 SW 1 n This page intentionally left blank ...
Page 137: ...Instructions for Use Oxylog 3000 SW 1 n 137 This page intentionally left blank ...