Getting Started
42
Instructions for Use Oxylog 3000 SW 1.n
Check connections
1
Ensure that the gas supply has been
connected.
2
Ensure that the correct gas type (O
2
or medical
air) has been selected.
If the wrong gas type has inadvertently been
set:
– Press the
Alarm Reset
key to cancel the
device check.
– Set correct gas type in configuration
Select
gas supply
and restart device check.
– Restart the device check.
3
Ensure that the test lung has been connected.
The Oxylog 3000 automatically checks if a test lung
has been connected. The device check is aborted
if a test lung is not detected within one minute.
The check is continued when the test lung is
detected.
4
Ensure that the configured hose system has
been connected, either
– the disposable hose system
or
– the reusable hose system.
5
Confirm the appropriate hose system. The
second page of the device check appears.
If the wrong hose system has been configured:
– Press the
Alarm Reset
key to cancel the
device check.
– Select the correct hose system. Refer to
"Select hose type" on page 81.
– Restart the device check.
System check
1
Set the controls (A) below the display to the
required values.
The Oxylog 3000 successively activates the
audible and visual alarm signals and prompts the
operator to acknowledge each signal.
2
Confirm the audible and visual alarm signals.
The device check continues automatically.
During the automatic test sequence, the Oxylog
3000 checks the flow, pressure levels and alarm
signals. Corresponding sounds are heard.
The bar graph shows the progress made by the
check.
The result is displayed on the Oxylog 3000
.
3
Confirm. The system returns to the menu
screen.
A monthly check of the power failure alarm is
recommended.
1
Switch the device on.
2
Disconnect the external power supply.
3
Remove the battery to activate the audible
alarm signal.
4
Listen for the audible alarm.
5
When the power failure alarm test is completed,
reinstall the battery into the battery compart-
ment of the Oxylog 3000.
023
NOTE:
Contact DrägerService if no alarm is heard.
Oxylog 3000
A
Summary of Contents for Oxylog 3000
Page 26: ...26 Instructions for Use Oxylog 3000 SW 1 n This page intentionally left blank ...
Page 44: ...44 Instructions for Use Oxylog 3000 SW 1 n This page intentionally left blank ...
Page 74: ...74 Instructions for Use Oxylog 3000 SW 1 n This page intentionally left blank ...
Page 86: ...86 Instructions for Use Oxylog 3000 SW 1 n This page intentionally left blank ...
Page 102: ...102 Instructions for Use Oxylog 3000 SW 1 n This page intentionally left blank ...
Page 106: ...106 Instructions for Use Oxylog 3000 SW 1 n This page intentionally left blank ...
Page 110: ...110 Instructions for Use Oxylog 3000 SW 1 n This page intentionally left blank ...
Page 134: ...134 Instructions for Use Oxylog 3000 SW 1 n This page intentionally left blank ...
Page 137: ...Instructions for Use Oxylog 3000 SW 1 n 137 This page intentionally left blank ...