Instructions for Use Oxylog 3000 SW 1.n
41
Getting Started
Perform device check
The device check consists of the following steps.
– Duration is approximately 3 minutes.
Connect the test lung
1
Connect the angled connector (A) to the
breathing valve.
2
Connect the catheter connector (B), diameter
7 mm, to the angled connector.
The catheter connector simulates the
resistance of the airways.
3
Connect the test lung (C).
Switch ON
1
To switch ON briefly press the
O
key (A).
The device performs a self-test and the operator is
prompted, on the display, to activate the configura-
tion menu or device check:
Press rotary knob for device check and
configuration
2
Press the rotary knob (B) to confirm, before the
bar is full.
3
Select
Device check
in the main menu and
confirm.
021
CAUTION
BTPS values of a test lung are not the same as the
BTPS values of a patient. The Oxylog 3000 mea-
sures and adapts according to BTPS values of a
patient. Therefore, when a test lung is connected,
the MVe and VTe indicated on display may differ
from the MVe and VTe that is set by the
operator.
A
C
B
022
119
37171
NOTE:
The device check can be discontinued at any time
by pressing the
Alarm Reset
key.
Oxylog 3000
A
B
Summary of Contents for Oxylog 3000
Page 26: ...26 Instructions for Use Oxylog 3000 SW 1 n This page intentionally left blank ...
Page 44: ...44 Instructions for Use Oxylog 3000 SW 1 n This page intentionally left blank ...
Page 74: ...74 Instructions for Use Oxylog 3000 SW 1 n This page intentionally left blank ...
Page 86: ...86 Instructions for Use Oxylog 3000 SW 1 n This page intentionally left blank ...
Page 102: ...102 Instructions for Use Oxylog 3000 SW 1 n This page intentionally left blank ...
Page 106: ...106 Instructions for Use Oxylog 3000 SW 1 n This page intentionally left blank ...
Page 110: ...110 Instructions for Use Oxylog 3000 SW 1 n This page intentionally left blank ...
Page 134: ...134 Instructions for Use Oxylog 3000 SW 1 n This page intentionally left blank ...
Page 137: ...Instructions for Use Oxylog 3000 SW 1 n 137 This page intentionally left blank ...