Alarms
66
Instructions for Use Oxylog 3000 SW 1.n
In the event of a gas failure
In the event of an internal power failure
CAUTION
In the event of a gas failure, proper ventilation
cannot be guaranteed. The Oxylog 3000 issues
the alarm
!! Supply pressure low
.
Immediately start ventilating the patient with an
independent manual ventilation device (resuscita-
tion bag) using PEEP and/or increased inspiratory
oxygen concentration where appropriate.
WARNING
In the event of an internal power failure, auto-
matic ventilation, volume measurement and
alarms do not operate!
An audible alarm announces the internal
power failure.
Spontaneous breathing can continue through
the emergency air intake.
Immediately start ventilating the patient with
an independent manual ventilation device
(resuscitation bag) using PEEP and/or
increased inspiratory oxygen concentration
where appropriate.
Summary of Contents for Oxylog 3000
Page 26: ...26 Instructions for Use Oxylog 3000 SW 1 n This page intentionally left blank ...
Page 44: ...44 Instructions for Use Oxylog 3000 SW 1 n This page intentionally left blank ...
Page 74: ...74 Instructions for Use Oxylog 3000 SW 1 n This page intentionally left blank ...
Page 86: ...86 Instructions for Use Oxylog 3000 SW 1 n This page intentionally left blank ...
Page 102: ...102 Instructions for Use Oxylog 3000 SW 1 n This page intentionally left blank ...
Page 106: ...106 Instructions for Use Oxylog 3000 SW 1 n This page intentionally left blank ...
Page 110: ...110 Instructions for Use Oxylog 3000 SW 1 n This page intentionally left blank ...
Page 134: ...134 Instructions for Use Oxylog 3000 SW 1 n This page intentionally left blank ...
Page 137: ...Instructions for Use Oxylog 3000 SW 1 n 137 This page intentionally left blank ...