
Set machine to »standby«
In »
standby
« mode,
– the system can immediately be switched to any
operating mode,
– gas consumption is zero,
– power consumption is marginal and
– the piston cylinder unit is in withdrawal position.
1
Press
A
on the ventilator for at least three
seconds.
2
Display
Standby
The flow tubes are no longer lit up.
The monitor remains operational and should be set to
»standby« by pressing
E
if not required.
The monitor can only be switched to standby if the
ventilator is already in standby!
During breaks:
●
Set fresh gas flow to »
O
«.
●
On Vapor: set control dial to »
0
«.
After use
If operation is interrupted for several hours, we
recommend that you do not leave the Cato in »Standby«
mode but switch it off completely.
●
Turn the master switch to »
O
« = Off –
the switch symbol is no longer white.
After switching off, an OFF delay displays the
message »
Power off
« in the Ventilator display for
about 10 seconds.
●
Disconnect the gas hoses from the wall sockets, or
close the cylinder valves of the gas cylinders.
●
Roll up hoses and hang them over the holder at the
back of the machine.
●
Disconnect the mains plug.
Operation
Set machine to »standby«
After use
48
1
2
erase
trend
Select ventilation mode at the ventilator
or switch on screen by pressing
Standby
Cato
SW-Version 2.05
02-05-00
config.
00337225
09128970