If the self-test reveals a
fault impairing the equipment's
safety
, the following message appears on the monitor:
»
Not ready
«.
The machine cannot switch to Standby and cannot be
switched to measuring mode. DrägerService must be
called immediately.
●
The configuration menu is displayed after the self-test
(example):
Fresh gas - External outlet
(optional)
Fresh gas outlet for connecting to semi-open systems.
1
Fit the hose of the semi-open system to the external
fresh gas outlet.
●
Make sure the anaesthetic gas scavenging hose is
connected.
2
Set the desired O
2
and N
2
O flow-rates on the
flowmeter block.
3
Press »
MAN SPONT
« on the ventilator.
4
The lamp in the »
FRESH GAS OUTLET
« key does not
light up.
●
Check that pressure is building up in the circle system.
4
Switch on the fresh gas outlet –
Press »
FRESH GAS OUTLET
« key.
The lamp in the key lights up.
●
Check that pressure builds up in the semi-open
system.
27
Preparing for use
Checking readiness for operation with checklist
Self-test
Fresh gas - External outlet
Standby / Configuration
✓
anaesth. gas
Halothane
Enflurane
Isoflurane
Sevoflurane
Desflurane
no
anaesth. gas
warning
default
✓
calibrating
O
2
-sensor
21 Vol.%
flow sensor
Ventilator
start up test
more
✓
Alarms inactive!
defaults
pulse to.
0 1 2 3 4
5
6 7 8 9
alarm tone
1 2 3
4
5 6 7 8 9
mode
adult
Neo.
parameters
record
interfaces
alarm limits
curves
basic configuration
D
PM
8050
0
I
GA
S
1
3
4
2
01328971
01237224