Uninterruptible power supply
(optional, see separate Instructions for Use)
Make sure that the status line is connected, so that the
uninterruptible power supply (UPS) is controlled by Cato.
In the event of power failure, the machine is powered
by the battery of the UPS, but the auxiliary mains
socket on the back of the Cato will be de-energised.
However, auxiliary equipment powered by the refrigerator
connector on the side of the Cato will still be powered.
1
Plug the power plug of the Cato into the socket of the
UPS.
2
Plug the power plug of the UPS into the mains
socket.
The UPS can supply the Cato and connected auxiliaries
with electrical energy for about 45 minutes. It is auto-
matically activated in the event of a power failure.
External equipment
See page 53 for the configuration of the interfaces.
Connection via the protocol interface
3
with data cable for printers with
serial interface
, e.g.
Desk-Jet (Hewlett-Packard)
or
e.g. the PM 8060 Vitara Patient Monitor with the
MEDIBUS protocol.
See page 53 for configuration.
Connection via the Dräger RS 232 C MEDIBUS
interface
e.g. to connect the PM 8060 Vitara Patient Monitor.
4
Connect with data cable.
●
The equipment plugs must be secured with the
screws provided.
Preparing for use
Connecting the equipment
Uninterruptible power supply
External equipment