in the breathing hoses
– Water traps must be fitted at the lowest point in the
hose and hang downwards.
– They must be checked regularly and drained if
necessary.
Note hygiene regulations –
risk of infection!
The hose system remains sealed. Container must be
replaced securely.
"Tips on reducing condensation“, see page 136.
Checking the soda lime
The soda lime turns purple from the bottom upwards
when saturated with CO
2
. To monitor CO
2
absorption,
the FiCO
2
value is displayed in "data screen" mode.
An alarm limit of 5 mmHg is set for monitoring CO
2
absorption.
The soda lime must be changed when two-thirds of the
charge has changed colour! The colour may also fade
again due to drying out after prolonged breaks.
Changing soda lime during operation
(see also page 91)
●
Prepare a replacement soda lime container.
●
Occlude the Vapor and stop N
2
O delivery.
●
Set ventilator to »
MAN/SPONT
« and the pressure
limiting valve to »
SPONT
«.
1
Briefly turn the soda lime container anticlockwise and
pull it downwards.
●
Insert the replacement soda lime container from
below and turn clockwise as far as possible.
●
Gas and machine settings may be cancelled.
●
Remove spent soda lime from the container.
Note hygiene regulations when dealing with
infectious patients – risk of infection!
45
Operation
Replace water separator
Checking the soda lime
[