
Continuous tone – volume must remain constant for
30 seconds.
If not, reconnect to mains and leave machine on for
24 hours so that battery can recharge. Repeat test!
●
Switch off machine. Press power switch again and
turn to
O
. Alarm goes out.
●
Reconnect to power supply.
Self-test
The self-test must be completed successfully in order to
establish readiness for operation!
Supplementary equipment must not be switched on until
after the self-test.
Start self-test:
●
Switch on machine: press power switch.
Test proceeds as described from page 31 onwards.
Calibrate flow sensor,
see page 30
Calibrate inspiratory O
2
sensor with
100% O
2
This is only necessary if sidestream O
2
measurement is
not used.
●
Remove O
2
sensor and place test adapter 68 01 349
on sensor.
●
Allow an O
2
flow of approx. 1 L/min to flow over the
O
2
sensor for approx. 2 minutes.
●
Select »
more
« with the rotary control.
●
Select »
O
2
sensor 100 Vol.%
« with the rotary control.
●
Display:
When the O
2
sensor has been flushed with O
2
for
approx. 2 minutes:
●
Press rotary control to confirm; calibration is then
started and continues automatically.
The clock icon is replaced by a tick (
✓
) when calibration
is complete.
●
Replace the O
2
sensor in its mount.
103
Check operational readiness
Self-test
Calibrate flow sensor
Calibrate inspiratory O
2
sensor
Standby / Configuration
Alarms inactive!
O
2
sensor
100 vol. %
linearity
test O
2
CO
2
sensor
Detach O
2
sensor and
flush with 100 vol. % O
2
.
calibrating
O
2
sensor
21 vol.%
flow sensor
Ventilator-
start up test
more
✓
✓
✓
✓
Calibrate O
2
sensor after
replacing sensor and after 24 hrs.
O
2
ca. 1 L/min
07228971
19428970