
Check operational readiness
Maintenance intervals
106
The machine and machine parts must be cleaned and disinfected before starting any maintenance work
and before returning to the manufacturer for repair.
"WaterLock" water trap
See separate Instructions for Use.
O
2
sensor
Replace if calibration is no longer possible or if the
message
FIO
2
INOP!
is displayed.
Disposal: see page 88.
Flow sensor
Replace if calibration is no longer possible or if the
message
FLOW INOP!
is displayed.
Can be incinerated at temperatures above 800
°
C
with low pollutant emissions.
Bacterial filter in the
secretion aspiration line
Replace after 14 days.
Disposal: as infectious special waste.
Can be incinerated at temperatures above 800
°
C
with low pollutant emissions.
Bacterial filter in
measured gas return line
Replace every 6 months.
Can be incinerated at temperatures above 800
°
C
with low pollutant emissions.
Bacterial filter in
pressure measuring line
Replace every 6 months.
Can be incinerated at temperatures above 800
°
C
with low pollutant emissions.
Cooling air filter
Clean monthly and dry well, or replace filter.
Replace after 1 year at the latest.
Disposal: as domestic waste.
Lithium battery for data storage
Must be replaced by qualified personnel after 2 years.
Disposal: see page 88.
Optical test bench for measuring
anaesthetic gas concentration
Must be inspected every 6 months by qualified personnel.
Time Keeper RAM
Must be replaced after 3 years by qualified personnel.
Disposal: in conformity with local waste disposal regulations.
Inspection and servicing
Every 6 months by qualified personnel.
Maintenance intervals