
Breathing system
Lift off table top:
1
Lever set to the position shown in black.
The following are complete:
2
Hose with manual ventilation bag
Symbol:
(Connected from below – not shown)
– Correct breathing hoses installed.
(Adult or infant hoses)
3
Pressure measuring hose with filter connected.
4
Measured gas return hose
Symbol:
connected.
5
Fresh gas hose plugged in.
(Connected from below – not shown)
6
Valve discs inserted.
– Pressure limiting valve (APL) present.
– Inspiratory microbial filter.
Symbol:
– Expiratory microbial filter.
Symbol:
Soda lime
7
Lime has not noticeably changed colour (purple).
– Filling level adequate (up to the mark).
– Soda lime container is securely tightened –
up to the
end stop
(clockwise).
Emergency ventilation bag
(not shown)
– Bag is complete with mask and hung from the side
of the Cato.
– Bag functions correctly.
✔
Check and tick off.
23
Preparing for use
Checking readiness for operation with checklist
Breathing system, Soda lime
Emergency ventilation bag
1
2
7
5
APL
3
4
6
03528970
03628970
03728970