Technical Data
Ambient conditions
For operation:
Temperature
20
°
C to 30
°
C (68
°
F to 104
°
F)
Atmospheric pressure
900 hPa to 1060 hPa
Relative humidity
0 to 75 %, no condensation
Air velocity in climate rooms
max. 0.3 m/s
For storage and transport:
Temperature
–20
°
C to 60
°
C (–4
°
F to 140
°
F)
Atmospheric pressure
500 hPa to 1060 hPa
Relative humidity
0 to 90 %, no condensation
Radiant heater
Output (irradiance) at 80 cm distance
from heater
Heater level 3
10 mW/cm
2
Heater level 10
30 mW/cm
2
Lights
120 V
23 W for work light, 7 W for night light
Minimum distance required between
top of radiant heater and ceiling
>50 cm (20 inches)
Skin temperature measurement
Sensor
Use only original Dräger sensors
(see "Ordering Information", page 65)
Measurement range
15
°
C to 42
°
C
Accuracy
±
0.1
°
C
Accuracy of entire measurement system
±
0.2
°
C
Setting range
35
°
C to 37.5
°
C
Mattress heater (Babytherm 8010 only)
Temperature measurement:
Measurement range
5
°
C to 45
°
C
Accuracy
±
0.5
°
C between 20 and 42 ºC
±
2 ºC otherwise
Setting range
30
°
C to 38.5
°
C
Height adjustment (available option)
Height adjustment range
295 mm
Specified operation
Intermittent operation
(Maximum operating time of 1 min must be followed
by 60 min period for cooling.)
Technical Data
61
Operating Instructions Babytherm 8004/8010, 2. ed.