1
Press
»OK«
key = confirm heater output at suggested
level.
Or
2
press
p
or
P
key to select heater output level.
1
Press
»OK«
key = confirm heater output.
NOTE:
If
»OK«
key is not pressed within approximately
10 seconds, the message will disappear and the previous
setting will remain in effect.
3
Display of newly selected heater output level.
4
Yellow LED in the key is now continuously lit,
" manual" mode is now active.
As a reminder for the required monitoring of core
temperature, an audible alarm is generated every
14 minutes at heater levels 4 or higher.
5
Yellow LED starts blinking, and
●
red central alarm light on the radiant heater starts
flashing.
6
display of set heater level starts blinking, and
7
text message appears:
15 min. patient temp check required.
Press OK-button to confirm.
NOTE:
After 15 minutes (one minute from time of
message), the heater would shut off if alarm not
acknowledged.
●
Check patient temperature.
8
Press
G
key, its yellow LED will be lit continuously.
Or
9
press
»OK«
key.
The audible alarm will stop, the red central alarm light will
go off, and the heater will be switched on again (will
remain switched on).
NOTE
: No reminder alarms are scheduled up to heater
level 3.
NOTE:
A skin temperature probe that might be
connected simultaneously in manual mode will provide a
skin temperature reading. It will, however, not control
heater output.
WARNING !
Measure core temperature at regular intervals!
Operation
Setting Heater Output Manually
29
Operating Instructions Babytherm 8004/8010, 2. ed.
15 min
Babytherm 8010
1
5
10
<36
°
C
°
C
skin
man.
skin
°
C
manual
mattress.
°
C
3
4
2
1
15 min
Babytherm 8010
1
5
10
<36
°
C
°
C
skin
man.
skin
°
C
manual
mattress.
°
C
5
8
9
7
6