Using Phototherapy (Available Option)
●
Swivel radiant unit over patient. Phototherapy lights
must be positioned vertically above patient for full
effectiveness of the therapy.
Switching phototherapy lights on
1
Press key, yellow LED in the key starts blinking.
2
Display shows message:
Phototherapy: Use eye protection
Press OK-button to start XX:XX:XX
.
XX:XX:XX
= time from the first activation of phototherapy
for the baby up to the switch off of the complete
Babytherm or up to the reset of the counter.
3
Press
»OK«
key = phototherapy lights are switched
on.
2
Display shows for 5 seconds:
Phototherapy duration
XX:XX:XX
1
Yellow LED in key is now continuously lit.
If radiant heater is operated in "manual" mode:
●
Reduce heater output approximately 3 levels down
from setting used while phototherapy lights were off.
NOTE:
Radiant heater output is adjusted automatically to
patient needs when operated in "Skin" mode.
To reset the babycounter phototherapy
1
Press key for 3 seconds,
2
Display shows:
Phototherapy duration
XX:XX:XX
Press OK to reset the counter.
3
Press
»OK«
key = reset,
3
Display shows:
Phototherapy duration
00:00:0X
Switching phototherapy lights off
1
Press key
2
Display shows message
Phototherapy OFF
Press OK-button to confirm
.
3
Press
»OK«
key = phototherapy lights switched off.
1
Yellow LED in key is now off.
WARNING !
Check the baby´s core temperature at regular
intervals. Phototherapy increases heat transfer to
patient.
Measure patient's bilirubin level at regular intervals.
WARNING !
Always consider differentiated diagnosis for patient
so life sustaining measures are not delayed.
WARNING !
Always use patient eye protection when using
phototherapy.
Operation
Using Phototherapy (Availble Option)
43
Operating Instructions Babytherm 8004/8010, 2. ed.
15 min
Babytherm 8010
1
5
10
<36
°
C
°
C
skin
man.
skin
°
C
manual
mattress.
°
C
3
1
2