Deviations greater than
±
0.5
°
C between set and
measured value for skin temperature:
1
Display of skin temperature starts blinking.
2
Yellow LED starts blinking.
Additionally, flashing red central alarm light on the
radiant heater.
3
Display shows message:
Skin temp. deviation above 0.5
°
C
Press OK-button to acknowledge
.
The alarm threshold for skin temperature deviations may
be adjusted in configuration mode, see page 45.
The factory set default is
±
0.5
°
C.
Audible alarm can be silenced and red central alarm light
be switched off for 10 minutes:
4
Press
G
key.
or
5
press
»OK«
key.
2
Yellow LED stays continuously lit.
●
Red central alarm light is switched off.
After skin temperature has returned to a value within
±
0.5
°
C of set temperature, the yellow LED, red central
alarm light, and audible alarm are turned off.
Prior to removing skin temperature probe temporarily,
switch to "manual" mode.
Refer to "Setting Heater Output Manually", page 28.
Skin temperature above 39
°
C
6
Display of skin temperature starts blinking.
7
Red LED starts blinking, together with audible alarm.
Additionally, blinking red central alarm light on the
radiant heater.
8
Display shows message:
Skin temperature above 39
°
C
Press OK-button to acknowledge
.
7
Red LED is continuously lit.
In "manual Mode" the heater is switched off at this time.
Audible alarm can be silenced and red central alarm light
be switched off for 2 minutes:
9
Press
G
key.
or
10
press
»OK«
key.
When skin temperature has returned to measured values
below 39
°
C, the alarm is automatically canceled.
Operation
Using Skin Temperature Control
34
Operating Instructions Babytherm 8004/8010, 2. ed.
15 min
Babytherm 8010
1
5
10
<36
°
C
°
C
skin
man.
skin
°
C
manual
mattress.
°
C
2
1
3
5
4
15 min
Babytherm 8010
1
5
10
<36
°
C
°
C
skin
man.
skin
°
C
manual
mattress.
°
C
8
7
9
6
10