Parameter
Option
1.
Averaging (AVg)
A1
1a.
Analog Filter
5
2.
Zero Tracking (AZt)
0.5
3.
Motion Aperture (MA)
1
4.
Serial Output (d.o.)
T.O.D.
5.
Startup Unit (S.U.)
LB
6.
Scale Capacity (CAP)
Variable
7.
Startup Zero (S.U.O)
AU0
8.
Operating Mode (oP)
Std
9.
Push Button Enable (P.b.)
non
9a.
Beep On
10.
Displayed Resolution (CtS) 3
11.
Tolerance Entry (toL)
S42
12.
Tare Entry (tAr)
Set
13.
Simplex Baud Rate (b.r.t)
120
13a.
Duplex Baud rate (d.b.r)
120
14.
Data Bits and Parity (d.b.P) 8 N
14a.
Duplex (d.b.1)
8 N
15.
Print Format (PFt)
1
16.
Checkweigh Operation (C.o.)
b A
17.
Convert Operation Sel. (CSL)
LGO
18.
Display Operation (doP)
on
19.
Handshaking (hSh)
non
20.
Unit On Time (tdY)
on
21.
Scale I.D. (i.d.)
01
22.
Setpoint Entry Mode (S.P.t) SET
23.
Setpoint Output Operation (S.P.o) HoF
24.
Time and Date Format (t.d.O)
non
25.
Print Line/Form Feed Select
non
Notes:
i.
Parameter 2 should be set to off or 0.5 divisions to comply with the approval.
ii.
Parameter 8 should be set up to STD to allow full range zero. Otherwise the zero
range will be limited.
iii.
Parameter 10 should always be set to 3 since the scale is only certified to 3000
divisions.
iv.
Parameter 16 should always be set to operate in both the Gross and Net modes
since the scale will normally be in the Gross mode only.
Section 9 Model 4100L