ENTERING THE SETUP MODE
To enter the Setup Mode, it is necessary to gain access to the main board. This is done
differently with the different 4200 models (see figure 1). Note: indicator must be
powered-up, before switching S2 - position 9 or S1 to the Setup Mode.
Model 4200 Tower: Remove the small rectangular access panel on the rear of the tower
by removing the two hex screws holding it in place. Flip S1 located on
the PCB just inside the access port (red-handled toggle
switch)
to enter the setup mode.
Model 4200M:
Remove the rectangular access panel on the rear of the indicator by
removing the six hex head screws holding it in place. Locate S2 -
position 9 using figure 2. Move this switch to the closed position to
enter the setup menu.
Model 4200RSS:
Loosen the front cover latches and open the front cover. Locate S1
on the main board using figure 2. Flip S1 to enter the setup mode.
When the switch has been closed, the scale display will respond by showing the message
"AUG --". You are now ready to begin programming the 4200.
SELECTING A CAPACITY AND RESOLUTION
After entering the setup mode as explained above, use the DOWN arrow button to advance
to the CAP parameter. Use the UP arrow button to advance to the desired capacity. Use
the DOWN arrow again to advance to the CtS parameter, and use the UP arrow to choose
3000 or 6000 divisions of displayed resolution (press the PRINT button to choose 10000
divisions).
ANALOG CALIBRATION
A.
If not already in the setup mode, do so now using the procedure outlined above.
B.
Press GROSS/NET to enter the calibration mode.
C.
Remove all weight from the platform.
D.
Press and hold the PRINT button to display the internal or raw counts - this is the
RAWLO value. The reading should be between 1000 and 10,000 counts.
E.
Place a test weight on the platform.
Section 2 Quick Setup Guide